's and Pacemakers — Transcript
Full transcript
- 0:04the objectives for this lecture are as
- 0:07follows describe indications for a ICD
- 0:10and pacemaker placement discuss the
- 0:12various modes of pacing identify
- 0:14potential complications with pacemakers
- 0:16and AI CDs during surgery and describe
- 0:19anesthetic management for patients with
- 0:21an AI CV or pacemaker this slide is
- 0:26adapted from the a sa practice advisory
- 0:28for perioperative management's of
- 0:29patients with a pacemaker or
- 0:31defibrillator and describes preoperative
- 0:33recommendations this slide is adapted
- 0:37from the a sa practice advisory for
- 0:39perioperative management of patients
- 0:41with the pacemaker or defibrillator and
- 0:42describes intraoperative recommendations
- 0:46this slide is adapted from the a sa
- 0:49practice advisory for perioperative
- 0:50management of patients with the
- 0:52pacemaker or defibrillator and describes
- 0:54post-operative recommendations
- 0:57battery-operated pacing devices were
- 1:00introduced by CW Lillehei a
- 1:02cardiothoracic surgeon and Earl Bakken
- 1:04and electrical technician in 1958 just
- 1:07four years after the invention of the
- 1:09transistor in 1960 Wilson Greatbatch an
- 1:12engineer in Buffalo New York
- 1:14created the first implantable
- 1:15battery-operated device in his barn the
- 1:18natural progression of pacemaker
- 1:19development led to invention of the
- 1:21implantable cardioverter-defibrillator
- 1:22around 1980 by Michael Moore cour of
- 1:25Baltimore first approved by the US Food
- 1:28and Drug Administration in 1985
- 1:30implantation of ICD s required a
- 1:32thoracotomy a significant operation in a
- 1:34patient with poor heart function
- 1:36advances and electronic miniaturization
- 1:39as well as improvements in battery
- 1:40technology have led to the development
- 1:42of very small but electronically
- 1:44complicated programmable pacing devices
- 1:49tecnológica ban says like transvenous
- 1:51lead placement anti tachycardia pacing
- 1:54capability miniaturization permitting
- 1:57pectoral pocket placement and multiple
- 1:59chamber and numerous survival benefits
- 2:01have led to increases in the uses of
- 2:03ICDs
- 2:04these changes in ICD implantation and
- 2:06indications have two important results
- 2:08for health care providers first a
- 2:11pectoral rather than abdominal pocket
- 2:13ICD web pacing capability might be
- 2:15mistaken by virtue of pacing spikes on
- 2:17the surface of an electrocardiogram for
- 2:19a non ICD pacemaker at many centers
- 2:23electrocardiograms are obtained from
- 2:24patients who have pacemakers with the
- 2:26use of a magnet because some I CDs from
- 2:29Boston Scientific guidance and CPI can
- 2:32undergo permanent deactivation of anti
- 2:34tachycardia therapy with magnetic
- 2:36placement this mistake could leave a
- 2:37patient unprotected and possibly result
- 2:39in death
- 2:40second pacing functions in an ICD often
- 2:43respond to external stimuli like
- 2:45magnetic placement or electromagnetic
- 2:47interferences differently than a
- 2:49pacemaker does these issues are
- 2:51addressed later in this lecture
- 2:53additionally all ICD is now perform
- 2:55permanent cardiac pacing so the issues
- 2:58in the section on pacing will apply to
- 2:59ICD care as well distinguishing a
- 3:02conventional pacemaker from an ICD can
- 3:04be accomplished by examining the right
- 3:06ventricular lead system on a chest
- 3:08radiograph a chest x-ray can also be
- 3:10used to identify the generator
- 3:12manufacture
- 3:15pacemaking and defibrillation systems
- 3:17have a remarkable record for reliability
- 3:19but they can and do fail myzel and
- 3:22colleagues reviewed FDA records from
- 3:24pacemakers and ICD failures either a
- 3:26death or explantation because of
- 3:28malfunction other than battery depletion
- 3:29from the periods of 1990 to 2002 even
- 3:34though the nature of this reporting
- 3:35scheme promotes under reporting of
- 3:36problems for these 12 years 2.25 million
- 3:39pacemakers were implanted with eight
- 3:42thousand eight hundred and thirty-four
- 3:43confirmed and reported malfunctions and
- 3:4630 deaths for ICD s four hundred and
- 3:49fifteen thousand seven hundred and
- 3:50eighty were implanted with eight
- 3:52thousand four hundred and eighty nine
- 3:53confirmed malfunctions and 31 deaths
- 3:56thus every patient with a cardiac
- 3:58generator should undergo regular
- 3:59in-office follow-up for appropriately
- 4:02selected devices telephone checks can be
- 4:04used to identify battery depletion of
- 4:06either a pacemaker or ICD or the
- 4:09detection of tachycardia but telephone
- 4:11checks cannot ensure adequacy of pacing
- 4:13capture and sent safety margins at this
- 4:15time for pacemakers nationally published
- 4:18guidelines recommend at least a
- 4:20telephone check every one to three
- 4:21months depending on implant duration no
- 4:24such guidelines exist for ICD follow-up
- 4:27although most ICD manufacturers
- 4:29recommend follow-up periods not to
- 4:30exceed four months the complexity of
- 4:35pacemakers and ICDs as well as a
- 4:37multitude of programmable parameters
- 4:39limits a number of generalizations that
- 4:41can be made about the perioperative care
- 4:42of a patient with an implanted pulse
- 4:44generator population ageing continued
- 4:47enhancements and implantable technology
- 4:49and new indications for implantation of
- 4:51cardiac devices continue to drive the
- 4:53growing number of implants practice
- 4:56guidelines have been issued by the
- 4:57American College of Cardiology here
- 4:59after called the ACC the American Heart
- 5:01Association the AAA and the heart rhythm
- 5:04society formerly the North American
- 5:06Society of pacing and electrophysiology
- 5:08reviews can also be found in a number of
- 5:11journals patients with an implant and
- 5:13cardiac pulse generator often have
- 5:15significant comorbid disease in addition
- 5:17to their cardiac rhythm disturbance our
- 5:19ability to care for these patients
- 5:21requires attention to both their medical
- 5:23and psychological problems we don't have
- 5:25to understand every aspect of the
- 5:27complex
- 5:28programming that goes into these devices
- 5:29but we do need a general understanding
- 5:31of their pulse generator its function
- 5:33and it's probable idiosyncrasies in the
- 5:36operating or procedure room finally not
- 5:39all electronic generators implanted in
- 5:40the chest are cardiac devices and
- 5:42devices resembling cardiac pulse
- 5:44generators are being implanted at
- 5:46increasing rates for indications
- 5:47unrelated to cardiac disease when
- 5:50implanted in the pectoral position the
- 5:52usual place for current cardiac
- 5:53generators these non cardiac devices can
- 5:56be mistakenly identified as cardiac
- 5:57generators pulse generator implantation
- 6:00has been approved by the FDA for pain
- 6:02control thalmic stimulation to control
- 6:05Parkinson's disease for an ignorant to
- 6:07simulate the diaphragm and paralyze
- 6:09patients and vagus nerve stimulation to
- 6:12control epilepsy and depression vagus
- 6:14nerve stimulation is also under
- 6:16consideration for the treatment of heart
- 6:17failure and perhaps obesity thus when
- 6:20evaluating a patient with any pulse
- 6:22generator one must now determine whether
- 6:24the pulse generator will be pacing the
- 6:26heart stimulating the central nervous
- 6:27system stimulating the spinal cord or
- 6:30stimulating the vagus nerve as a side
- 6:33note although the device is commonly
- 6:34referred to as an automatic implantable
- 6:36cardioverter-defibrillator or a ICD the
- 6:39term a ICD is a brand name originally
- 6:41owned by CPI or cardiac pacemakers
- 6:44incorporated which was acquired by
- 6:46guidance Medical Corporation and is now
- 6:48part of Boston Scientific the term PCD
- 6:51for programmable
- 6:52cardioverter-defibrillator belongs to
- 6:54Medtronic corporation let's look at
- 6:58pacemakers pacemaker manufacturers
- 7:01report that more than 2,500 generators
- 7:03have been manufactured by more than
- 7:04twenty-six named entities over the past
- 7:0640 years currently more than 250,000
- 7:10adults and children in the United States
- 7:12undergo new pacemaker placement each
- 7:13year and nearly 2 million patients have
- 7:16pacemakers today many factors lead to
- 7:19confusion regarding the behavior of a
- 7:20device and the perioperative care of a
- 7:22patient with the device especially since
- 7:24case reports text books and literature
- 7:26reviews have not kept pace with
- 7:27technological developments a pace making
- 7:31system consists of an impulse generator
- 7:33and a lead or leads to cure the
- 7:36electrical impulse to the patient's
- 7:37heart leads are connected to the heart's
- 7:39chambers through the vena cava
- 7:42user transvenous leads or are directly
- 7:44sewn onto the surface of the heart and
- 7:46they are called epicardial leads leads
- 7:48can be unipolar one electro / lead
- 7:51bipolar two electrodes / lead or
- 7:54multipolar multiple electrodes and wires
- 7:57contained within one leaf with
- 7:58connections in multiple chambers because
- 8:01two electrodes are required to complete
- 8:03a circuit the second electrode in a
- 8:05unipolar configuration will be the metal
- 8:07generator case use of the case as an
- 8:09electrode requires that the generator
- 8:11pocket be devoid of gas and electrical
- 8:13continuity has reportedly been disrupted
- 8:15by the use of nitrous oxide pacemakers
- 8:20with unipolar leads seem to be more
- 8:21sensitive to the effects of EMI or
- 8:23electromagnetic interference and these
- 8:26systems produce a larger spike on an
- 8:28analogue recorded electrocardiogram most
- 8:31pacemaker systems pace in bipolar mode
- 8:34because bipolar pacing usually requires
- 8:36less energy
- 8:37bipolar sensing is more resistant to
- 8:39interference from muscle artifacts or
- 8:41stray electromagnetic fields frequently
- 8:44bipolar electrodes can be identified on
- 8:46the chest x-ray because they will have a
- 8:48ring electrode 1 to 3 centimeters
- 8:50proximal to the lead tip note however
- 8:53that generators what bipolar leads can
- 8:55be programmed to the unipolar mode for
- 8:57pacing sensing or both no discussion of
- 9:02pacemakers can take place without an
- 9:03understanding of the generic pacemaker
- 9:05code which has been published by the
- 9:07NASPE
- 9:08and British pacing and electrophysiology
- 9:11group this code called nbg was initially
- 9:14published in 1983 and was last revised
- 9:17in February of 2002 and describes the
- 9:19basic behavior of the pacing device the
- 9:22first two positions of this code
- 9:24chambers paste and chamber sense seems
- 9:26relatively straightforward although
- 9:28early pacemakers provided only
- 9:30ventricular support current models can
- 9:32provide pacing in the atria and
- 9:33ventricles and these devices can also be
- 9:36programmed to determine intrinsic
- 9:37activity in these chambers as well the
- 9:40code does not describe the array of
- 9:42diagnostic data that can be accumulated
- 9:43by these devices probably the most
- 9:46confusing aspect of the nbg code is the
- 9:49third position which is the response to
- 9:51the sensed event most pacemakers are
- 9:54programmed either to the DD
- 9:55mode dual-chamber pacing instancing both
- 9:58triggered and inhibited mode or the vvi
- 10:01mode for single chamber ventricular
- 10:03pacing in the inhibited mode two other
- 10:06modes frequently found are VDD
- 10:08ventricular pacing with dual chamber
- 10:11pacing in the triggered and inhibited
- 10:12mode and DDI dual chamber pacing and
- 10:16sensing but inhibited mode only this
- 10:20third position describes the following
- 10:22behavior dual inhibited or triggered DDD
- 10:26pacing provides atrioventricular or AV
- 10:29synchrony in the DDD setting atrial
- 10:32pacing will take place in the inhibited
- 10:34mode that is the pacing device will emit
- 10:37an atrial pulse if no sensed atrial
- 10:39event takes place within the appropriate
- 10:41time frame in the DDD or VDD devices
- 10:45once an atrial event has occurred
- 10:47whether native or paced the pacing
- 10:49device will ensure that a ventricular
- 10:51event follows an inhibited the
- 10:54appropriate chamber is paced unless
- 10:55intrinsic electrical activity is
- 10:57detected during the pacing interval for
- 10:59the DDI mode AV synchrony is provided
- 11:02only when the atrium is paste if
- 11:04intrinsic atrial activity is present no
- 11:07AV synchrony is provided by the pacing
- 11:09function in triggered the pacing device
- 11:12will emit a pulse only if a response to
- 11:14a sensed event the triggered mode is
- 11:16used when the device is being tested
- 11:20the modes V DD and DD I deserve further
- 11:24comment V DD pacing is used for a
- 11:27patient with AV nodal dysfunction but
- 11:29intact an appropriate sinus node
- 11:31behavior V DD pacing is accomplished
- 11:34with a single lead that incorporates
- 11:36atrial sensing electrodes and
- 11:37ventricular conductors that can both
- 11:39pace and sense a V DD device has no
- 11:43atrial pacing capability as a result in
- 11:46a patient who depends on atrial
- 11:48contraction to augment cardiac output
- 11:50events that result in V VI pacing like a
- 11:53sinus rate below the programmed rate or
- 11:54battery depletion or asynchronous
- 11:57ventricular pacing can lead to
- 11:58deteriorating hemodynamics d di pacing
- 12:02is indicated for a patient who has a
- 12:03dual chamber pacing device and also has
- 12:06episodes of paroxysmal atrial
- 12:07dysrhythmias d di pacing prevents high
- 12:11ventricular rates that could result from
- 12:12attempted tracking of the atrial
- 12:14arrhythmia and it provides AV synchrony
- 12:16only when the atrium is paced many ddd
- 12:20programs will enter the DDI mode on
- 12:22detection of high atrial rates called
- 12:24mode switch automatic mode switch or
- 12:27atrial tachyons depending on the
- 12:29manufacturer when the mode switches to
- 12:32DDI perturbations such as a very high
- 12:35atrial rate or EMI might revert the
- 12:37pacemaker back to DDD pacing rate
- 12:42modulation the fourth position also
- 12:44remains a poorly understood concept
- 12:46because some patients cannot increase
- 12:48their heart rate in response to
- 12:50increased oxygen demand pacemaker
- 12:52manufacturers have devised a number of
- 12:53mechanisms to detect patient exercise
- 12:55such as sensors that detect vibration
- 12:58respiration and pressure as the sensor
- 13:00detects exercise it increases the pacing
- 13:03rate termed sensor indicated rate as the
- 13:06exercise tapers this sensor indicated
- 13:08rate returns to the program to lower
- 13:10rate the sensitivity of these sensors to
- 13:12their exercise signals and the rates of
- 13:14change in pacing are programmable
- 13:16features and current generators
- 13:18activation of rate response algorithms
- 13:20resulting in an increased pacing rate in
- 13:23the operating room whether from vigorous
- 13:25chest wall skin preparation pressure on
- 13:27the generator or EMI has led to
- 13:29inappropriate treatment and patient harm
- 13:31with the 2002 revision
- 13:34the nbg the fifth column describes
- 13:36multi-site pacing functionality it has
- 13:39been used to describe anti tachycardia
- 13:41functions but this scheme has been
- 13:43abandoned and a generic defibrillator
- 13:45code has been established atrial
- 13:47multi-site pacing might prevent atrial
- 13:49fibrillation and ventricular multi-site
- 13:52pacing is an acceptable means of pacing
- 13:54patients with dilated cardiomyopathy
- 13:58there are many indications for permanent
- 14:00pacing classically anti bradycardia
- 14:03pacing treats patient with sinus node
- 14:05disease or improper impulse formation an
- 14:08AV node disease which is improper
- 14:10impulse conduction pacing can be used to
- 14:13reduce the outflow tract obstruction in
- 14:15hypertrophic obstructive cardiomyopathy
- 14:16in both adults and children because
- 14:19paced ventricular conduction takes place
- 14:21in a left bundle branch pattern meaning
- 14:23the left ventricular septum depolarizes
- 14:25after the other segments rather than as
- 14:27an early systolic event finally in
- 14:30August of 2001 devices were approved by
- 14:33the FDA for three chamber pacing right
- 14:36atrium and both ventricles to treat DCM
- 14:38this pacing modality can also be called
- 14:41cardiac resynchronization therapy or
- 14:43biventricular pacing to accomplish LV
- 14:47pacing a pacing lead is placed in the
- 14:49coronary sinus or sewn onto the LV free
- 14:51wall pacing for hokum and DCM requires
- 14:55careful attention to pacer programming
- 14:57to be effective in these patients
- 14:59the PMS provides a stimulus for
- 15:01ventricular depolarization and AV
- 15:03synchrony must be preserved p.m.
- 15:06inhibition or loss of pacing from native
- 15:09conduction atrial irregularity
- 15:11ventricular irregularity the development
- 15:13of a junctional rhythm or EMI can lead
- 15:16to deteriorating hemodynamics in these
- 15:18patients by be pacing might cause
- 15:21inappropriate lengthening of the QT
- 15:23interval in susceptible patients and
- 15:24this lengthening has been reported to be
- 15:26associated with torsades as a result of
- 15:29this report a prudent Anessa too should
- 15:31ensure adequate access to rapid
- 15:33defibrillation for patients with by be
- 15:35pacing accomplished without an ICD
- 15:40despite oft-repeated folklore most
- 15:43pacemaker manufacturers warn that
- 15:45magnets were never intended to treat
- 15:46pacemaker emergencies or prevent the
- 15:48effects of EMI rather magnet activated
- 15:52switches were incorporated to produce
- 15:53pacing behavior that demonstrates
- 15:55remaining battery life and sometimes
- 15:57pacing threshold safety factors
- 15:59placement of a magnet over a generator
- 16:01might produce no change in pacing
- 16:03because not all pacemakers switched to a
- 16:06continuous asynchronous mode when a
- 16:08magnet is applied moreover not all
- 16:11models from a given company behave the
- 16:12same way in some devices magnet behavior
- 16:15can be altered by programming whereas in
- 16:18others magnet behavior can be completely
- 16:19eliminated by programming
- 16:23for all generators calling the
- 16:26manufacturer remains the most reliable
- 16:28method of determining magnet response
- 16:30and using this response to predict
- 16:31remaining battery life as battery
- 16:34voltage Falls the magnet response can be
- 16:36used to detect the following IFI
- 16:38intensified follow-up required the
- 16:41device must be checked frequently
- 16:42approximately every four weeks for most
- 16:44models er I elect the replacement
- 16:48indicator the device is nearing the end
- 16:50of its useful life and should be
- 16:51electively replaced
- 16:52Yoel end of life the device has
- 16:55insufficient battery power remaining and
- 16:57should be replaced immediately pre
- 17:01anesthetic management of a patient with
- 17:03a pacemaker includes the evaluation and
- 17:05optimization of coexisting diseases ACC
- 17:09guidelines suggest that cardiac testing
- 17:10be dictated by the patient's underlying
- 17:12disease medications symptomology
- 17:15interval from last testing and planned
- 17:17intervention no special laboratory tests
- 17:20or radiographs are needed for patients
- 17:22with a conventional pacemaker chest
- 17:24x-ray is rarely the pick leave problems
- 17:26and not all devices have radiographic
- 17:28markings however a patient with a baby
- 17:31pacemaker or ICD might need a chest film
- 17:34to document the position of the coronary
- 17:35sinus lead especially if central line
- 17:37placement is planned currently there are
- 17:40no data or guidelines regarding
- 17:41placement of a central line in the
- 17:43setting of a coronary sinus lis and
- 17:45because these leads have no fixation
- 17:47they might be more easily dislodged than
- 17:49the standard pacemaker or ICD lead in
- 17:51fact in early studies spontaneous
- 17:54dislodgement of the coronary sinus lead
- 17:56was detected in more than eleven percent
- 17:57of patients appropriate reprogramming is
- 18:01the safest way to avoid intra operative
- 18:02problems especially if mono polar or
- 18:05Bovie electro surgery will be used many
- 18:08pacemaker manufacturers stand ready to
- 18:09assist with this task however
- 18:12interrogation of any generator with or
- 18:14without reprogramming requires a medical
- 18:16prescription and supervision by an
- 18:18appropriately trained physician hospital
- 18:20guidelines and policy should reflect the
- 18:22need for a competent physician to review
- 18:24and validate any care delivered by a
- 18:26company representative although no
- 18:30special monitoring or anaesthetic
- 18:32technique is required for a patient with
- 18:33a cardiac generator attention must be
- 18:36given to a number of
- 18:37first ECG monitoring of a patient must
- 18:40include the ability to detect pacing
- 18:42discharges
- 18:43currently most ECG monitors in both the
- 18:46operating room and the intensive care
- 18:48unit perform digital acquisition and
- 18:50analysis of electrocardiographic signals
- 18:53and they are subject to considerable
- 18:54interference from a variety of sources
- 18:56in their default settings these monitors
- 18:59often filter the pacing artifacts and no
- 19:02pacing spikes will be shown this
- 19:04filtering must be disabled so that the
- 19:06monitor will paint pacing spikes onto
- 19:08the display
- 19:09even with the filtering disabled however
- 19:11pacing artifacts do not always appear
- 19:13atrial signals are weak and current
- 19:16monitors do a poor job of detecting them
- 19:18moreover because many of the digital
- 19:20monitors analyze only one lead for these
- 19:22signals and then paint on artifacts on
- 19:25every lead placement of the ECG leads
- 19:27can markedly affect the detection access
- 19:29in practice when a patient with known
- 19:32pacing is monitored sometimes changing
- 19:34the analysis lead on the monitor will
- 19:36result in the appearance of pacing
- 19:38signals unfortunately when the pacemaker
- 19:40artifact filter is disabled
- 19:42EMI especially from the use of mono
- 19:45polar es you can lead to inappropriate
- 19:47painting of pacing artifacts on the
- 19:49monitor second patient monitoring must
- 19:52include the ability to ensure that the
- 19:54pace electrical activity is converted to
- 19:56mechanical cicely mechanical Cicely is
- 19:59best evaluated by pulse oximetry plasma
- 20:02graphi or arterial pressure waveform
- 20:04display
- 20:07third there remains limited published
- 20:10perioperative experience with
- 20:12intraoperative IV pacing at this time
- 20:14these patients often have ejection
- 20:16fractions of less than 30 percent and
- 20:18they depend on pacing in both ventricles
- 20:20to improve their cardiac output loss of
- 20:23ventricular pacing from any cause like a
- 20:25VD synchrony atrial fibrillation or
- 20:27flutter appearance of a junctional
- 20:29rhythm myocardial ischemia an acid-base
- 20:32disturbance or a changing in the pacing
- 20:35threshold can cause an immediate
- 20:36decrease in cardiac output with the
- 20:39exception of transesophageal
- 20:40echocardiography no b2b monitoring of
- 20:43cardiac output has demonstrated any
- 20:45utility in detecting loss of by be
- 20:47pacing behavior patients with hocum
- 20:50pacing can be dependent on ventricular
- 20:52pacing to limit LV outflow tract
- 20:54obstruction fourth some patients might
- 20:57need an increase pacing rate during the
- 20:58perioperative period to meet the
- 21:00increased oxygen demand this subject is
- 21:02often not addressed pacemaker patients
- 21:05reportedly experience post-operative
- 21:07morbidity and mortality frequently and
- 21:09failure to address tissue oxygen demands
- 21:11and cardiac output needs might
- 21:13contribute to this problem this
- 21:15appropriate equipment must be on hand to
- 21:17provide backup pacing or defibrillation
- 21:19or both if needed
- 21:21cardiac generators although handy
- 21:23occasionally performs some untoward
- 21:25maneuver or fail even in the absence of
- 21:27EMI acceptable but inappropriate
- 21:30behavior of a pacemaker or ICD can
- 21:32create an inhospitable situation even a
- 21:36properly working dual-chamber pacemaker
- 21:37can produce rnt pacing especially in the
- 21:40setting of junctional rhythm or
- 21:42premature ventricular contractions
- 21:44the medical team caring for the patient
- 21:46with an implanted cardiac pulse
- 21:47generator must understand that the
- 21:49patient has been deemed needy of this
- 21:51device by a physician who is an expert
- 21:52in the diagnosis and management of
- 21:54cardiac rhythm issues very few of us are
- 21:57qualified to contradict this diagnosis
- 21:59yet some persists in providing an
- 22:01anesthetic without appropriate backup
- 22:02pacing and deeper relation equipment on
- 22:04hand
- 22:09mono polar Bo ves you use remains the
- 22:12principal enter operative issue for
- 22:14patients with a pacemaker between 1984
- 22:17and 1997 the FDA was notified of 456
- 22:21adverse events with pulse generators 255
- 22:24from electro surgery and a significant
- 22:26number of device failures a mono polar
- 22:29ESU is more likely to cause problems
- 22:31than a bipolar ESU and pacemakers with a
- 22:34unipolar electrode configuration are
- 22:36more sensitive to EMI than those with
- 22:38bipolar configurations the most common
- 22:41effect of an e su on pacing function is
- 22:44ventricular / sensing which causes
- 22:46pacing inhibition sometimes the
- 22:48generator detects significant EMI and
- 22:50begins pacing asynchronously at the
- 22:52programmed lower rate this behavior is
- 22:55called noise reversion mode pacing even
- 22:58though the pacemaker does not actually
- 22:59change modes noise reversion is not
- 23:02present in some I CDs and is
- 23:04programmable in others if a mono polar
- 23:09ESU is to be used the electrosurgical
- 23:11current return pad often misidentified
- 23:13as a grounding pad must be placed to
- 23:16ensure that the e su current path does
- 23:18not cross the pacemaking system some
- 23:21authors recommend placement of this pad
- 23:23on the shoulder for head and neck
- 23:24procedures or on the distal part of the
- 23:26arm with a sterile drape of the wire for
- 23:28breast and axillary procedures
- 23:32procedures using only a mono polar ESU
- 23:35or with special emi pacing ramifications
- 23:38include the following lithotripsy
- 23:40transurethral resection and uterine
- 23:43hysteroscopy magnetic resonance imaging
- 23:45electroconvulsive therapy nerve
- 23:48simulator testing or therapy or
- 23:50succinylcholine or etomidate use most
- 23:54monitors manufactured since 2005 have
- 23:56much improved capability to detect and
- 23:58display pacing artifacts although they
- 24:00still filter these high-frequency
- 24:01signals in their default setup mode
- 24:05pacemaker failure has three causes
- 24:07generator failure lead failure or
- 24:10failure of capture generator failure is
- 24:13rare and a device that has previously
- 24:15been evaluated and not near the end of
- 24:17useful battery life unless the generator
- 24:19or leads is struck directly by the ESU
- 24:22bleep failure also unusual but reported
- 24:25during patient repositioning can result
- 24:27in under sensing / sensing or failure to
- 24:30deliver sufficient energy to the
- 24:31myocardium to produce depolarization
- 24:34term loss of capture myocardial changes
- 24:37that lengthen the refractory period or
- 24:39increasing energy requirements for
- 24:41depolarization can result from
- 24:42myocardial ischemia or infarction
- 24:44acid-base disturbance electrolyte
- 24:47abnormalities or abnormal
- 24:49anti-arrhythmic drug levels the response
- 24:53to pacemaker failure depends on the
- 24:55clinical situation a patient with a
- 24:57perfusing rhythm and stable vital signs
- 25:00can be observed while a plan is made to
- 25:01correct the problem for a patient with
- 25:04inadequate profusion the following steps
- 25:06can be attempted
- 25:07while cardiopulmonary resuscitation is
- 25:09in progress when appropriate one a
- 25:11magnet can be applied if the pacemaker
- 25:13is known to revert to an asynchronous
- 25:15mode which will also eliminate sensing
- 25:17behavior in these devices for many of
- 25:20the st. Jude auto capture devices with
- 25:22magnet mode enabled magnet application
- 25:24might increase pacing amplitude and
- 25:26restore capture devices from ela medical
- 25:30typically increase their pacing
- 25:31amplitude when a magnet is applied some
- 25:34pacemakers perform threshold margin
- 25:36testing upon magnet placement which can
- 25:38further reduce pacing to temporary
- 25:41pacing can be initiated and it can be
- 25:42transfer assic transcutaneous
- 25:44transvenous or transesophageal
- 25:47transesophageal atrial pacing requires a
- 25:50functional atrium and AV node for
- 25:52ventricular activation so it is
- 25:54contraindicated in patients with atrial
- 25:55fibrillation or flutter it is also
- 25:58generally contraindicated in the
- 26:00presence of a permanent pacemaker or ICD
- 26:02in the setting of external pacing the
- 26:05electrocardiogram can be misinterpreted
- 26:06because the pacemaker artifacts are
- 26:08large in comparison to the QRS complexes
- 26:12successful ventricular pacing has been
- 26:14reported with transesophageal pacemakers
- 26:16but it remains unreliable and produces
- 26:18inferior hemodynamics when compared with
- 26:20atrial pacing note that any external
- 26:23pacing might further inhibit pacemaker
- 26:25output at energies that will not produce
- 26:27myocardial capture 3 sympathy mimetic
- 26:30drugs can decrease the polarization
- 26:32threshold or increase chrono tropa city
- 26:35or both
- 26:36epinephrine 0.5 to 1 mic per minute or
- 26:39dopamine 5 to 20 mics per kilo per
- 26:42minute should be considered
- 26:43isoproterenol 0.5 mics per minute is
- 26:47often recommended but not widely
- 26:48available and is complicated by
- 26:50hypotension antimuscarinic drugs like
- 26:53atropine and glycol pie relate might be
- 26:55helpful for causes of myocardial
- 26:58ischemia should be sought and corrected
- 27:00myocardial ischemia can substantially
- 27:02increase the energy required for
- 27:04ventricular capture 5 disturbances and
- 27:07electrolyte balance anti-arrhythmic drug
- 27:09levels and acid-base equilibrium should
- 27:11be investigated and corrected potassium
- 27:14calcium and magnesium abnormalities can
- 27:17raise depolarization thresholds in
- 27:19addition potassium flux ionized calcium
- 27:22levels and acid-base equilibrium can be
- 27:25affected by hyperventilation and
- 27:27hypoventilation 6 if none of the
- 27:30measures just presented succeed
- 27:31consideration should be given to the
- 27:33placement of an epicardial lead by the
- 27:35surgical staff a pacemaker that was
- 27:39reprogrammed for the perioperative
- 27:40period should be reset appropriately for
- 27:43non reprogrammed devices most
- 27:45manufacturers recommend interrogation to
- 27:47ensure proper functioning and acceptable
- 27:49remaining battery life if any electro
- 27:51surgery was used the ACC guidelines now
- 27:54recommend a post procedure interrogation
- 27:57consideration should be given to
- 27:58increasing the lower paced rate in a
- 28:00chrono tropically incompetent patient if
- 28:02increase cardiac output will benefit the
- 28:04patient during convalescence the
- 28:08development of an implantable
- 28:10battery-powered device able to deliver
- 28:12sufficient energy to terminate v-tach or
- 28:14ventricular fibrillation represents a
- 28:16major medical breakthrough for patients
- 28:18with a history of ventricular
- 28:19tachyrhythmia these devices reduce
- 28:22deaths in the setting of malignant
- 28:23ventricular tachyrhythmia x' and they
- 28:25clearly remain superior to
- 28:27anti-arrhythmic drug therapy initially
- 28:30approved by the fda in 1985 more than
- 28:3260,000 devices will be implanted in the
- 28:34United States this year an industry
- 28:36sources report that more than 250,000
- 28:39patients have these devices today
- 28:41furthermore results from the sudden
- 28:43cardiac death heart failure trial have
- 28:45led to a significant increase in the
- 28:47number of patients for whom ICD therapy
- 28:49is
- 28:49indicated a considerable number of
- 28:52technological visas have been applied
- 28:53since the first ICD was placed including
- 28:56substantial miniaturization as well as
- 28:58battery improvements that now permit
- 29:00permanent pacing with these devices thus
- 29:03one could easily confuse a pectoral ICD
- 29:05for a pacemaker like pacemakers I see
- 29:08these have a four place to generic code
- 29:10to indicate lead placement and function
- 29:12the most robust form of identification
- 29:14called the label form experience the
- 29:17fourth character into its component
- 29:18generic pacemaker code newer icds since
- 29:231993 have many programmable features but
- 29:26essentially they measure each cardiac
- 29:28Arta are interval and categorize the
- 29:30rate as normal too fast or too slow when
- 29:34the device detects a sufficient number
- 29:35of short r2r intervals within a certain
- 29:38period it will begin an anti tachycardia
- 29:40event the internal computer will choose
- 29:43ATP or shock depending on the
- 29:45manifestation and device programming if
- 29:48shock is chosen an internal capacitor is
- 29:50charged charging time is dependent on
- 29:53the desired output and it can be from
- 29:55six to fifteen seconds for a maximum
- 29:57shock charging time is lengthened by low
- 29:59battery voltage time from last charge
- 30:01and low temperature most IC these can be
- 30:06programmed to reconfirm v-tach or v-fib
- 30:09after charging to prevent inappropriate
- 30:11shock therapy some IC DS can be
- 30:14programmed to begin ATP immediately on
- 30:16initiating a charge cycle typically I
- 30:19see these deliver six to eighteen shocks
- 30:21per event once the shock is delivered no
- 30:24further ATP will take place despite
- 30:27considerable improvement in the
- 30:28detection of ventricular dysrhythmias
- 30:29more than 10% of shocks are for rhythms
- 30:32other than v-tach or v-fib
- 30:34super ventricular tachycardia remains
- 30:36the most common etiology of
- 30:38inappropriate shock therapy and causes
- 30:40of inappropriate shock have been
- 30:41reviewed elsewhere programmable features
- 30:44and current IC DS to differentiate
- 30:46v-tach from a tachycardia of super
- 30:48ventricular origin include the following
- 30:50onset criteria
- 30:52stability criteria QRS with criteria the
- 30:56intelligence and the morphology waveform
- 30:59analysis
- 31:01an ICD with antibiotic Ardea Therapy
- 31:05capability will begin pacing when the
- 31:07RTR interval is too long
- 31:09in July of 1997 the FDA approved devices
- 31:12with sophisticated dual chamber pacing
- 31:14modes and rate response behavior for ICD
- 31:17patients who need permanent pacing which
- 31:19is about 20% of ICD patients because of
- 31:22the David study which suggested that the
- 31:25DDD pacing in a patient with a clear
- 31:27need for a dual chamber pacing decreases
- 31:29survival when compared with a single
- 31:31chamber device placement many
- 31:33electrophysiologists our programming
- 31:34long AV delays in these patients to
- 31:36limit ventricular pacing as noted
- 31:39previously however long a V delays can
- 31:42lead to our on T pacing in addition many
- 31:45ICD manufacturers have designed
- 31:47algorithms to limit ventricular pacing
- 31:49some of these algorithms allow dropped
- 31:51QRS events which can resemble a
- 31:54second-degree mobitz to block or
- 31:55malfunction of the pacing system
- 31:59initially I see these were placed for
- 32:01hemodynamically significant v-tach or
- 32:03v-fib newer indications associated with
- 32:06sudden death include patients awaiting
- 32:08heart transplantation and those with
- 32:10long QT syndrome Brugada syndrome and a
- 32:13red magenta RB dysplasia studies suggest
- 32:17that icds can be placed prophylactically
- 32:19to prevent sudden death in young
- 32:20patients with hypertrophic
- 32:21cardiomyopathy as well as in post
- 32:24myocardial infarction patients with an
- 32:26ejection fraction less than 30 percent
- 32:27but they are no benefit after coronary
- 32:30artery bypass grafting like pacemakers
- 32:34magnet behavior and ICDs can be altered
- 32:37by programming most devices will suspend
- 32:40tacky dysrhythmia detection and
- 32:42therefore therapy when a magnet is
- 32:44appropriately placed to activate the
- 32:45magnet switch some devices from ng on
- 32:48Boston Scientific CPI guidance
- 32:52pacesetter
- 32:53st. Jude medical or venture Tech's can
- 32:56be programmed to ignore magnet placement
- 32:58and some guidance icds have their magnet
- 33:00function permanently disabled because of
- 33:02a magnet switch issue depending on
- 33:05programming
- 33:05anteye tachycardia therapy and some
- 33:07Boston Scientific guided medical and CPI
- 33:10devices can be permanently disabled by
- 33:12magnet placement for 30 seconds and as
- 33:15earlier some patients have been
- 33:16discovered with their ICD anti
- 33:18tachycardia therapy unintentionally
- 33:20disabled in boston scientific guidance
- 33:25and CPI devices if the magnet mode is
- 33:28enabled and the ICD is enabled for anti
- 33:31tachycardia therapy the ICD will emit
- 33:33beep synchronized to our waves to
- 33:35signify adequate placement of the magnet
- 33:37and suspension of the tachyarrhythmia
- 33:39detection hence disabling of the therapy
- 33:42if the ICD emits a constant tone anti
- 33:45tachycardia therapy has been programmed
- 33:47to off regardless of magnet placement
- 33:50depending on programming ICD is from
- 33:53boston scientific guidance medical and
- 33:55CPI can toggle between activated and
- 33:57inactivated states with 30 seconds of
- 33:59magnet application thus to re-enable
- 34:02therapy the magnet must be removed and
- 34:05then replaced until the constant tone
- 34:07reverts to beep synchronized to our
- 34:08waves subsequent removal of the magnet
- 34:11then returns the anti tachycardia
- 34:13therapy to the enabled state any Boston
- 34:16Scientific Gaiden or CPI device that
- 34:19emits either beeps or constant tones
- 34:21while a magnet is in place will have its
- 34:23anti tachycardia therapy disabled during
- 34:25the magnet session in general magnets
- 34:30will not affect ICD anti bradycardia
- 34:33pacing modes or rates enter medics
- 34:36devices transiently change the pacing
- 34:38rate to reflect battery voltage again
- 34:41interrogating the device and calling the
- 34:43manufacturer remain the most reliable
- 34:45method for determining magnet response
- 34:48in addition to evaluating and optimizing
- 34:52any comorbid disease and ICD patients it
- 34:54is often suggested that every ICD should
- 34:57undergo preoperative interrogation these
- 34:59devices store considerable data
- 35:01regarding the occurrence of this read
- 35:03Mia's because ATP is well tolerated most
- 35:06patients are not aware of this
- 35:07intervention for any patients scheduled
- 35:10to undergo an elective procedure the
- 35:12onset of a new dysrhythmia probably
- 35:14warrants investigation of the problem
- 35:16before the procedure here is a case
- 35:18study where unexpected ventricular
- 35:20tachycardia with anti tachycardia pacing
- 35:22was found during a visit from a surgeon
- 35:24requesting cardiac clearance a 65
- 35:27year-old woman with a history of beat
- 35:29had undergone implantation of a
- 35:30Medtronic single-chamber defibrillator
- 35:32about eight months previously she had
- 35:35not had any dizziness or syncopal
- 35:37episode since the placement
- 35:39interrogation of her device in the
- 35:40preoperative center revealed vve v VI
- 35:43programing along with an episode of
- 35:45tachycardia at 150 to 160 beats per
- 35:48minute that was detected by the ICD s
- 35:51v-tach
- 35:51the ICD delivered a 6-feet burst of anti
- 35:54tachycardia pacing at 182 beats per
- 35:57minute which converted the tachycardia
- 35:59back to sinus rhythm no backup pacing
- 36:02was needed after the v-tach was
- 36:03terminated this device was set to detect
- 36:06v-tach at 16 consecutive ventricular
- 36:08events with a rate between 146 and 200
- 36:11per minute and to deliver anti
- 36:13tachycardia pacing at 84% of the last
- 36:16RTR interval the surgery was
- 36:19consequently postponed pending further
- 36:20evaluation the determination of the need
- 36:25for elective replacement of an ICD
- 36:27because of battery depletion is more
- 36:29complicated than with pacemakers because
- 36:32some ICDs can have multiple battery
- 36:34cells predicting battery depletion based
- 36:36on battery voltage is difficult in
- 36:38general however the manufacturer should
- 36:40be consulted for any device with a
- 36:42charging time in excess of 12 seconds
- 36:45most ICD should have their anti
- 36:47tachycardia therapy disabled before the
- 36:49use of any device that causes EMI the
- 36:52use of a mono polar es you can produce
- 36:54inappropriate shocks many icds have no
- 36:57noise reversion behavior so es un deuce
- 37:00ventricular / sensing might lead to non
- 37:02pacing in a patient who is dependent on
- 37:04the ICD for pacing at this time no
- 37:10special monitoring because of the ICD is
- 37:12required for a patient with an ICD
- 37:16electrocardiographic monitoring and the
- 37:17ability to deliver external
- 37:19cardioversion or defibrillation must be
- 37:21present during the time of ICD
- 37:23disablement should cardioversion or
- 37:26defibrillation be needed the
- 37:27defibrillator pads should be placed so
- 37:29that they avoid the pulse generator to
- 37:30the greatest extent possible
- 37:32nevertheless one should remember that
- 37:34the patient not the ICD is being treated
- 37:36the other recommendations that we
- 37:39discussed for the intraoperative
- 37:40management of pacemakers apply here as
- 37:42well
- 37:44no special anesthetic techniques have
- 37:46been championed for patients with an ICD
- 37:49most of these patients will have
- 37:50severely depressed systolic function
- 37:52dilated ventricular cavities and
- 37:54significant valvular regurgitation
- 37:57thus the anesthetic of choice should be
- 37:59dictated by the underlying physiological
- 38:01Arrangements that are present
- 38:03conflicting data have been published
- 38:04regarding the choice of an anesthetic
- 38:06agent or agents and changes in
- 38:08defibrillation threshold in 1993 Gill
- 38:11and co-workers examine defibrillation
- 38:13threshold and dogs and concluded that
- 38:15neither halothane nor isoflurane changed
- 38:18the DFT and open chest defibrillation
- 38:20when compared with pentobarbital and
- 38:22fusion however whine Burum and
- 38:25associates evaluated the defibrillation
- 38:27threshold in humans during ICD
- 38:29implantation and found that halothane
- 38:31isoflurane and fentanyl increased the
- 38:33DFT even with these increases the
- 38:36increased dfts were still substantially
- 38:38lower than the maximum energy generally
- 38:40available in ICDs
- 38:42and these increases would not have been
- 38:44noted under the usual testing conditions
- 38:45as a post anesthesia evaluation the ICD
- 38:51must be rien tera gated and re-enabled
- 38:53and deaths from failure to re-enable an
- 38:55ICD have been reported all recorded
- 38:58event should be reviewed encounter
- 39:00should be cleared the pacing parameters
- 39:02must be checked and reprogrammed as
- 39:04necessary in conclusion electronic
- 39:09miniaturization has permitted the design
- 39:11and the use of sophisticated electronics
- 39:13and patients who have need for
- 39:14artificial pacing or automated
- 39:16cardioversion defibrillation of their
- 39:18heart these devices are no longer
- 39:21confined to merely keeping the heart
- 39:22beating between a minimum and maximum
- 39:24rate they are now being used as therapy
- 39:26to improve a failing heart both aging of
- 39:29the population and our ability to care
- 39:31for a patient with increasingly complex
- 39:33disease suggest that we will be caring
- 39:35for many more patients with these
- 39:36devices and we must be prepared for this
- 39:38situation safe and efficient clinical
- 39:41management of these patients depends on
- 39:43our understanding of implantable systems
- 39:45there are indications of use and the
- 39:47perioperative needs that they create
- 39:48this slide is a reference a pulse
- 39:51generator companies for you companies
- 39:53that are bolded market implantable
- 39:55cardioverter to
- 39:56or bleeders
- 40:04you
About this transcript
This page contains the full transcript of 's and Pacemakers by Sterling Anesthesia Education Solutions, Inc., generated from the public captions YouTube serves with the video. The transcript has 5,949 words across 1,048 segments, with the original timestamps preserved so you can click any line to jump to that moment in the embedded player.
What you can do with it
Use the transcript to take notes, quote the speaker, build a study guide, generate a summary with ChatGPT or Claude via the YouTube Summary tool, or export it as a timed subtitle file with YouTube to SRT. You can also re-open it in the transcriber to translate the transcript into 100+ languages.
Free YouTube transcript tool
YouTube2Text is a free YouTube transcript generator — no signup, no daily limit. Paste any YouTube link and get the full transcript instantly, with timestamps, click-to-jump, translation to 100+ languages, AI prompts for ChatGPT, Claude, and Gemini, and exports to TXT, SRT, VTT, or Markdown.