Najgroźniejszy błąd w historii radioterapii — Transcript
Full transcript
- 0:00The patient felt a rush of heat, as he
- 0:02later recalled, as if someone had
- 0:04poured hot coffee on his back. The
- 0:06machine stopped working and displayed
- 0:08the message "malfunction 54." At the
- 0:10same time, the operator received
- 0:12information that the device had
- 0:14delivered a fraction of the required
- 0:16dose. Following procedure, she pressed
- 0:19"P" for "proceed" to continue. She
- 0:22didn’t know that in less than a
- 0:24second, the patient had received a
- 0:25radiation dose over 100 times higher
- 0:27than planned. And this was even before
- 0:30she pressed "B." We are talking about a
- 0:32radiotherapy device intended to destroy
- 0:34cancer cells. This story became one of
- 0:37the most infamous disasters at the
- 0:39intersection of medicine, physics,
- 0:40radiation, and computer science. My
- 0:42name is Karol Bielski. This is "
- 0:44Odwrócona Lekcja"(Inverted Lesson). If
- 0:46you feel like supporting us with a
- 0:47subscription, we would be very pleased.
- 0:49Though, of course, the decision is
- 0:51yours. Let’s get started. Chapter one
- 0:53. A new generation machine. In the
- 1:06early 70s, the Canadian state-owned
- 1:08company Atomic Energy of Canada Limited
- 1:10(AECL) collaborated with the French
- 1:12company CGR to produce radiotherapy
- 1:14equipment. The fruit of this
- 1:17cooperation was two modern machines for
- 1:19those times. The Therac-6 and the
- 1:22Therac-20. Both devices were equipped
- 1:25with a PDP-11 computer, but its role
- 1:27was marginal. It helped the operator
- 1:31enter parameters and simplified
- 1:32operation, while the machine’s most
- 1:34important functions could run
- 1:35independently of this computer. Safety
- 1:39was overseen by mechanical interlocks
- 1:40and separate electrical circuits. If
- 1:44individual parts of the device were not
- 1:46in the correct position during patient
- 1:48irradiation, the machine would not
- 1:49allow it. It was physically impossible
- 1:53to deliver the electron beam. In the
- 1:55mid-70s, AECL began work on the next
- 1:58generation of accelerator. The new
- 2:02design used a dual-pass electron system
- 2:04, allowing the device to achieve high
- 2:06energy while taking up less space. The
- 2:11first prototype was created in 1976,
- 2:14and at the end of 1982, the fully
- 2:16computerized version, the Therac-25,
- 2:18hit the market. Before we go any
- 2:23further, during this past vacation, I
- 2:25was resting while my money kept working
- 2:27, all thanks to Trade Republic. The
- 2:30partner of today's episode. Not long
- 2:33ago, I kept my money in several
- 2:34different accounts without a specific
- 2:36plan and, frankly, without much profit.
- 2:39Then I discovered Trade Republic and a
- 2:41lot has changed. It is a licensed
- 2:44German bank trusted by over 10 million
- 2:46customers in Europe. In one simple app,
- 2:50I save, pay, and even—albeit timidly
- 2:53for now—invest. What appealed to me
- 2:56most was the interest rate on cash.
- 2:58Trade Republic currently offers 6%per
- 3:00annum on funds in the account. Interest
- 3:03is calculated daily and paid out
- 3:05monthly, and the money is available at
- 3:07any time without locking it up. Unlike
- 3:10a fixed-term deposit, new customers
- 3:12receive 6%interest on balances up to
- 3:14200,000 PLN, and amounts above this
- 3:16limit are subject to interest equal to
- 3:19the NBP deposit rate. Here is an
- 3:21example. If you keep 10,000 PLN in the
- 3:23account, then at 6%per annum, this
- 3:25gives you 600 PLN gross per year.
- 3:28that's fifty gross per month, and you
- 3:30can see it all clearly right there in
- 3:32the app. These transparent rules are
- 3:35simply excellent. You open an account,
- 3:37turn on the interest in the app, and
- 3:39watch your earnings grow. With every
- 3:41interest payout, the withholding tax is
- 3:43automatically deducted, so you don't
- 3:45even have to worry about that. It's
- 3:48also worth mentioning that when paying
- 3:51abroad, there are no additional bank
- 3:53fees for currency conversion, and Trade
- 3:55Republic charges no fees for ATM
- 3:57withdrawals over 200 PLN. This was
- 4:00really useful for me during my holiday
- 4:02travels. Instead of looking for
- 4:04exchange offices, you just pay with
- 4:05your Trade Republic card and don't
- 4:07worry about anything else. Opening an
- 4:09account takes a few minutes and is done
- 4:11entirely online. Scan the QR code
- 4:13visible on the screen, download the app
- 4:16, and start saving. Starting today with
- 4:18Trade Republic. And now, let's get back
- 4:20to the material. In the first mode, the
- 4:22device directed an electron beam at the
- 4:25patient with energy from 5 to 25 MeV.
- 4:29Such a beam was primarily effective for
- 4:30treating lesions located relatively
- 4:32shallowly in the patient's body. The
- 4:35second mode was used for treating
- 4:36tumors located deeper. The Therac-25
- 4:39then accelerated electrons to an energy
- 4:41of 25 megaelectronvolts or more and
- 4:43directed them at a metal target. This
- 4:47collision led to the creation of X-ray
- 4:49radiation. X-ray radiation, which then
- 4:52passed through a special filter. To
- 4:56generate X-rays, an electron stream
- 4:58with an intensity about 100 times
- 5:00greater than during direct electron
- 5:02therapy was required. Inside the head,
- 5:06there was a rotating mechanism that
- 5:08positioned the elements we mentioned
- 5:10earlier, the target and the filter, in
- 5:12the path of the beam. The first
- 5:14position held the scanning magnets,
- 5:16used during electron therapy, the
- 5:17lighter one, let's say. The second held
- 5:21the target, the filter, and the chamber
- 5:23measuring the X-rays. The third
- 5:26position was used for aiming at the
- 5:27patient's body. In that position, a
- 5:31steel mirror was in the beam's path,
- 5:33and a simple light showed the operator
- 5:35the spot to be irradiated. In this last
- 5:39position, no dose-measuring sensor was
- 5:41installed. After all, the designers
- 5:44didn't anticipate that the accelerator
- 5:46could emit radiation then. I remind you
- 5:49, there was supposed to be only a light
- 5:51, you know, like a laser. A laser, not
- 5:54a light, which would simply show on the
- 5:55patient's body where we're going to
- 5:57shoot. And then the configuration would
- 6:02change to the first or second,
- 6:03depending on whether we want that
- 6:05lighter dose of electrons, or that
- 6:06X-ray radiation. If the device had
- 6:13triggered the strong beam intended for
- 6:15X-ray production, and the target and
- 6:17filter weren't in its path, the patient
- 6:19would have been hit by a concentrated
- 6:21stream of electrons without the
- 6:22elements meant to convert, weaken, and
- 6:24spread it. In older devices,
- 6:30independent mechanical and electrical
- 6:31interlocks protected against such a
- 6:33situation. When designing the Therac-25
- 6:37, AECL abandoned such old-fashioned and
- 6:39outdated safeguards. The manufacturer
- 6:44decided that a computer could
- 6:45simultaneously operate the device,
- 6:47check the position of the elements, and
- 6:49prevent the beam from starting in the
- 6:51wrong configuration. So, the
- 6:53responsibility was shifted from
- 6:55physical mechanical safeguards to
- 6:57digital locks. managed by a computer.
- 6:59The control program for the Therac-25
- 7:01was written in assembly for the PDP-11
- 7:04computer. It was developed by a single
- 7:06person. The entire software for a
- 7:09device that shoots radiation, electrons
- 7:11, at patients. One person, and they
- 7:16used code snippets that came from even
- 7:18older models. Documentation was sparse,
- 7:21and separate tests for the software
- 7:23itself were very limited in scope. The
- 7:27manufacturer later claimed during
- 7:29hearings that the system had undergone
- 7:31about 2,700 hours of testing. It turned
- 7:34out, however, that this number
- 7:36represented 2,700 hours of device usage
- 7:38. So, the tests were, well, they were
- 7:40only on the bodies of patients. In 1983
- 7:43, AECL conducted a safety analysis of
- 7:46the Therac-25. It assumed that
- 7:50simulator tests and prior usage had
- 7:51limited the number of programming
- 7:53errors. Perfect. Potential problems
- 7:57that might have remained in the code
- 7:59were not considered. The analysis
- 8:02covered possible computer failures, but
- 8:04it viewed them primarily as the result
- 8:06of hardware damage or random
- 8:08interference. The possibility that the
- 8:11program itself could have flaws was
- 8:13ignored. I got a bit confused. It was
- 8:16simply assumed that since everything
- 8:18worked for 2,000 hours, it would keep
- 8:20working, so why dig deeper? Ultimately,
- 8:2311 Therac-25 units were installed in
- 8:25the United States and Canada. The
- 8:29devices began treating patients and
- 8:30worked wonderfully for the first few
- 8:32years, and no officially confirmed
- 8:34problems were reported. Everything
- 8:38turns upside down on June 3, 1985, in
- 8:41Marietta, Georgia. After the beam was
- 8:45activated, the patient felt a wave of
- 8:47heat. When the operator entered the
- 8:50room, the woman said the machine had
- 8:51burned her. Please don't worry, that's
- 8:54impossible. Chapter two. That's
- 8:57impossible. The 61-year-old woman had
- 9:08previously undergone surgery to remove
- 9:10a breast tumor. Radiotherapy was
- 9:12intended to destroy cancer cells that
- 9:13might have remained in the surrounding
- 9:15lymph nodes. On June 3, '85, she was
- 9:19being prepared for another radiation
- 9:21treatment in the collarbone area with a
- 9:2310 MeV electron beam. Soon, redness and
- 9:28swelling appeared in the irradiated
- 9:30area. The pain grew, the shoulder
- 9:32stiffened, and muscles began to
- 9:34contract in such painful spasms. The
- 9:37woman ends up in the hospital, but
- 9:39doctors still referred her for further
- 9:40treatments at the center. Initially, it
- 9:44was suspected that the changes resulted
- 9:46from the progression of the disease or
- 9:48some unusual reaction to properly
- 9:49performed radiotherapy. However,
- 9:52medical physicist Tim Steel, who worked
- 9:54at the center, began to suspect a
- 9:56malfunction. He contacted AECL and
- 9:59asked if the Therac-25 could trigger an
- 10:01electron beam without the scanning
- 10:03magnets, which would, as it were,
- 10:05spread or disperse it across the
- 10:06treatment field. Such a situation, my
- 10:11dear sir, is. impossible. About two
- 10:14weeks after the procedure, redness
- 10:16appeared on the patient's back,
- 10:17corresponding to the lesions visible on
- 10:19the front of her body. The skin in the
- 10:21irradiated area begins to peel. The arm
- 10:23remains immobile. The woman suffers
- 10:26from constant pain. The nature of the
- 10:29injuries indicated severe radiation
- 10:31burns, but no one could explain how the
- 10:33Therac-25 could have caused them.
- 10:37Physicist Ken Kenstone later estimated
- 10:39that the patient may have received a
- 10:41dose of 15,000 to 20,000. At that time,
- 10:45they were measuring in rads, 15,000 to
- 10:4720,000 rads. A typical single
- 10:50therapeutic dose is about 200 rads. A
- 10:55precise reconstruction of the event
- 10:56proved impossible because the treatment
- 10:58parameter printing function was
- 11:00disabled at the time. There was no
- 11:04record left of the device settings.
- 11:07Meaning, no one knew exactly how the
- 11:09operator had set everything up to get
- 11:11that result. The woman's injuries were
- 11:13irreversible. Her breast was removed,
- 11:15she lost the ability to move her arm,
- 11:17and she was left with chronic pain. The
- 11:20other facilities—I remind you there
- 11:22were 11 of them—using the Therac-25
- 11:24were not informed about anything. The
- 11:28matter was also not reported to the
- 11:29American Food and Drug Administration.
- 11:32Regulations at the time imposed a duty
- 11:34to report malfunctions and injuries.
- 11:37Primarily on equipment manufacturers
- 11:39and importers. Hospitals and employees
- 11:42had no such obligation. At least not
- 11:44yet, because those standards were
- 11:46changed later. It is seven weeks after
- 11:48the accident in Marietta. We are moving
- 11:50to Canada now. On July 26, 1985, a 40-
- 11:53year-old patient arrives at the Ontario
- 11:56Cancer Foundation in Hamilton for her
- 11:5824th cervical cancer radiotherapy
- 12:00treatment. The operator starts the
- 12:05device, but after 5 seconds, the Therac
- 12:07-25 stops and displays an "H-tilt"
- 12:09error message, indicating beam
- 12:11asymmetry. Meaning something was wrong
- 12:14with the beam. However, the measurement
- 12:16system indicated that the patient was
- 12:18receiving no dose at all. The device
- 12:21classified the event as a pause, a
- 12:23break in operation, after which the
- 12:25treatment could be continued without
- 12:27re-entering parameters. So we have
- 12:31everything saved there, just press "P",
- 12:33no need to set anything up. You can
- 12:37fire electrons again. So the operator
- 12:39presses "P"—Proceed. Proceed. What's
- 12:43going on? It's not working again. "P"
- 12:45—The procedure is repeated. Damn it,
- 12:47it's jammed again. After the fifth
- 12:50attempt, the fifth attempt, the device
- 12:52went into a full halt state, as if it
- 12:53had frozen. For the staff, this was not
- 12:56an exceptional event. The Therac-25
- 12:59frequently interrupted work and
- 13:01displayed error messages about
- 13:02insufficient dose or incorrect beam
- 13:04setting. Usually, it was not associated
- 13:07with any danger to the patient. In some
- 13:09facilities, similar errors occurred
- 13:11even a dozen or several dozen times a
- 13:13day. In any case, five attempts were
- 13:16made. After the treatment, the patient
- 13:19complained of a burning sensation and a
- 13:21feeling resembling an electric shock in
- 13:23the hip area. Despite this, on the same
- 13:28day, six more people were irradiated
- 13:30using the same device. Three days later
- 13:33, the woman returns to the center. The
- 13:35pain would not subside. Swelling
- 13:37appears at the site of the treatment.
- 13:39The Therac-25 was taken out of service.
- 13:42The patient was admitted to the
- 13:43hospital with a suspected radiation
- 13:45overdose. AECL and Canadian radiation
- 13:48protection authorities were notified.
- 13:51Later estimates indicated the woman may
- 13:53have received 13,000 to 17,000 rads,
- 13:55though it is unknown how many of those
- 13:57five attempts were actual hits. Maybe
- 13:59all of them, or maybe just one, a very
- 14:01powerful one—it’s unknown. The
- 14:04patient died on November 3, 1985,
- 14:06though the autopsy stated it was due to
- 14:08an aggressive tumor, so the cause of
- 14:10death was cancer, not radiation.
- 14:16However, the injuries caused by the
- 14:17Therac were so severe that if she had
- 14:19survived, if the cancer hadn't killed
- 14:21her, she likely would have needed a
- 14:23total hip replacement. AECL engineers
- 14:27were unable to reproduce the failure.
- 14:29Suspicion fell on one of the
- 14:31microswitches that informed the
- 14:32computer about the position of the
- 14:34rotating mechanism. You know, it’s
- 14:35about that rotating mechanism that,
- 14:37when necessary, positioned the shield
- 14:39and the filter, or a mirror, or nothing
- 14:40at all. It was concluded that a
- 14:43momentary sensor failure could have
- 14:44sent incorrect information to the
- 14:46system and allowed the device to
- 14:47operate in the wrong configuration.
- 14:50AECL modified the sensor system and
- 14:52introduced additional software checks
- 14:54for the mechanism's position. The
- 14:57manufacturer claimed these changes
- 14:58reduced the risk of failure by at least
- 15:00five orders of magnitude. Something
- 15:03like 100,000 times less chance of a
- 15:05failure. Canadian experts recommended
- 15:08installing an independent
- 15:10hardware-based physical system to
- 15:11monitor the position of the rotating
- 15:13mechanism. They demanded that errors
- 15:17related to dose measurement immediately
- 15:19stop the treatment and prevent it from
- 15:21being resumed by just pressing a key.
- 15:23Meaning if something hangs, if
- 15:25something happens, if a message pops up
- 15:27, it shouldn't be enough to just press
- 15:29'P' to resume work. AECL did not
- 15:31implement all of those recommendations
- 15:33at the time. The number of possible
- 15:35retries was reduced from five to three
- 15:37before a system hang, but no
- 15:38independent additional lock for the
- 15:40mechanism's position was installed. A
- 15:44letter sent to users informed them of
- 15:45the issues with the device and
- 15:47recommended visually checking its
- 15:48settings. It did not mention that one
- 15:52of the patients had been seriously
- 15:53injured. After the modifications were
- 15:56made, the centers received permission
- 15:58to return to normal use of the Therac-
- 16:0025. Great. The manufacturer still
- 16:03didn't know what had happened. Chapter
- 16:05three: X and E. One of the devices
- 16:17modified after the accident in Hamilton
- 16:19was a Therac-25 operating at the Yakima
- 16:21Valley Memorial Hospital in Washington
- 16:23state. In December 1985, a patient
- 16:25treated there noticed red, stripe-like
- 16:28marks on her right hip. The staff tried
- 16:30to determine the cause. They suspected
- 16:33the elements used to shape the
- 16:34radiation field. They also considered
- 16:37reactions to chemotherapy. However,
- 16:39that should have appeared in other
- 16:40radiated areas as well and likely
- 16:42wouldn't have formed parallel stripes.
- 16:45When the staff learned that the woman
- 16:47used a heating pad at home, suspicion
- 16:49fell on the internal wires, where I
- 16:51think warm water flows. The pad was
- 16:53X-rayed, but the layout of the wires
- 16:55did not match the marks on the
- 16:57patient's body. So, the woman continued
- 16:59her radiotherapy. It ended on January 6
- 17:02, 1986. The staff still didn't know
- 17:05what caused the stripes, or these
- 17:06unusual injuries. On January 31,
- 17:08hospital staff described the case in a
- 17:11letter to AECL. They spoke by phone
- 17:13with a technical support manager. The
- 17:16response arrived on February 24, after
- 17:18almost a month, and AECL stated that
- 17:20the bodily injuries could not have been
- 17:22caused by either a Therac failure or
- 17:24operator error. They just couldn't
- 17:27imagine how such stripes could have
- 17:29been formed. They attached two pages of
- 17:33technical arguments as proof, claiming
- 17:35that such an operation, such a failure,
- 17:37was impossible. Furthermore, they
- 17:41argued that similar injuries had not
- 17:43been observed in other patients. Except
- 17:46we already know that while the injuries
- 17:48were not the same, they had been
- 17:49observed; I remind you of the cases
- 17:51from Kenstone and Hamilton. However,
- 17:52individual hospitals do not know that
- 17:54an accident is happening in another
- 17:56city somewhere. Each center seemed to
- 17:59only see its own piece of the problem,
- 18:01and this problem was much, much wider.
- 18:03The staff in Yakima trusted the
- 18:05manufacturer. The Therac had undergone
- 18:06modifications, and these were supposed
- 18:08to radically increase safety. I remind
- 18:10you, 100,000 times. Ultimately, the
- 18:13cause of the injuries was labeled as
- 18:15unknown. Less than a month after the
- 18:17response sent to Yakima, on March 21,
- 18:191986, a patient at the East Texas
- 18:21Cancer Center in Tyler arrived for his
- 18:24ninth radiotherapy treatment. He had
- 18:26recently had a tumor removed from his
- 18:27back, and the planned treatment was
- 18:29intended to destroy the remaining
- 18:31cancer cells. The doctor prescribed an
- 18:33electron beam with an energy of 22 MeV
- 18:35and a dose of 180 rads. The man lay
- 18:37down on his stomach. The operator left
- 18:39the room and closed the heavy door
- 18:41behind her. It is the kind of door that
- 18:43is supposed to protect her from
- 18:44radiation. After all, she spends the
- 18:46whole day somewhere near this machine.
- 18:48That day, the video monitor was
- 18:49disconnected, and the intercom was not
- 18:51working. They couldn't hear each other.
- 18:53That is, the female operator in one
- 18:55room and the patient behind the heavy
- 18:57door. The woman had been working with
- 19:00the Therac for a long time and operated
- 19:02the machinery and its keyboard very
- 19:03efficiently. While entering the
- 19:06parameters, out of habit, she chose the
- 19:08X-ray mode marked with the letter X,
- 19:09but she was a professional. Almost
- 19:12immediately, she noticed the mistake.
- 19:14She moved the cursor back and changed X
- 19:16from X to E, that is, from X-ray
- 19:18radiation to this electron radiation,
- 19:20those pure electrons. The other
- 19:23parameters matched, everything was
- 19:25correct, the operator confirmed it, and
- 19:27the computer displayed information that
- 19:29the device was ready for operation. So,
- 19:31she pressed the key to activate the
- 19:33beam. The patient felt a sudden strike
- 19:36and heat. He also heard a loud buzzing
- 19:39coming from the device. The Therac
- 19:41interrupted its operation; the screen
- 19:43displayed the message "Malfunction 54."
- 19:45The system indicated that instead of
- 19:47the planned dose, it had delivered a
- 19:49small fraction of it. The operator
- 19:51therefore followed the standard
- 19:53procedure, what the manufacturer
- 19:55recommended, and pressed "P" to proceed
- 19:57. At the same moment, the patient was
- 19:59trying to get up from the table. Well,
- 20:01they couldn't hear each other. He
- 20:03didn't know that she wanted to continue
- 20:04. Chapter four. Malfunction 54. The
- 20:16Therac activated the beam. The man felt
- 20:18a strike, this time resembling an
- 20:20electric shock running through his
- 20:21shoulder. He was already getting up. He
- 20:23was shot with the beam while he was in
- 20:25that position, you know, already
- 20:26half-sitting. He had the impression as
- 20:28if his hand had detached from the rest
- 20:30of his body. He got off the table,
- 20:32approached the door, and started
- 20:33banging on it. The operator immediately
- 20:35stopped the procedure and let him into
- 20:37the room next door, into the control
- 20:38room. The doctor noticed intense
- 20:40redness, a classic symptom, but assumed
- 20:42the man might have been shocked by
- 20:44electricity. Go home, and if anything
- 20:48happens, you know where to find us. A
- 20:51medical physicist working at the center
- 20:53, Fritz Heiger, was also called in. He
- 20:56checked the Therac’s calibration,
- 20:57finding nothing, no irregularities at
- 20:59all. Curiously enough, no one really
- 21:02knew what the "Malfunction 54" message
- 21:04meant. The next day, two AECL engineers
- 21:09arrived. One from the local branch, the
- 21:11other from headquarters in Canada. They
- 21:13tested the device all day, but they
- 21:15couldn't trigger the error. One of the
- 21:18engineers assured them that the Therac
- 21:2025 could not deliver an overdose. When
- 21:24Heiger asked about other cases, he was
- 21:26told that AECL knew of nothing similar.
- 21:30The manufacturer suggested the patient
- 21:32had indeed been shocked by electricity,
- 21:34so an independent engineering firm
- 21:36examined the installation but found no
- 21:38electrical fault. On April 7, the
- 21:40Therac returns to service. Four days
- 21:42later, on the 11th, another patient is
- 21:44treated. The same operator enters the
- 21:46treatment parameters. Damn, X instead
- 21:49of E again. Will I ever learn? She
- 21:51instantly notices the mistake, hits the
- 21:53correct key, changes the mode to
- 21:55electron, and quickly confirms the
- 21:57remaining data. The computer stated
- 21:59that the beam was ready. The operator
- 22:01activates the device. After a few
- 22:03seconds, the Therac stops with a buzz,
- 22:05displaying "Malfunction 54" on the
- 22:07screen. This time, the intercom was
- 22:10working. The operator heard moaning and
- 22:12immediately ran into the room. The man
- 22:14was trying to tear off the tape holding
- 22:16his head in place. He said he felt fire
- 22:17on the side of his face. I don’t
- 22:19think I mentioned that it was aimed
- 22:20right at his face. He later described
- 22:22that something had hit him. He saw a
- 22:25flash and heard a sound like eggs
- 22:27frying. This time, the device was
- 22:30immediately taken out of service. The
- 22:33operator remembered the sequence of
- 22:34actions perfectly. Accidentally
- 22:37selecting X, returning the cursor to
- 22:38the previous field, changing to E,
- 22:40confirming, and triggering the beam.
- 22:43Heiger tried to repeat the sequence,
- 22:45but nothing happened. The Therac was
- 22:48working correctly, and then it hit him:
- 22:50the speed. A physicist who didn’t
- 22:53operate the machine daily wasn’t as
- 22:55efficient; you know, he was like a
- 22:57woodpecker with a keyboard. He wasn't
- 22:59giving commands to the machine any
- 23:01slower than the experienced operator.
- 23:03He had to practice, and he actually
- 23:05learned to enter the corrections just
- 23:07as fast as she did. Once he succeeded,
- 23:10he immediately saw "Malfunction 54" on
- 23:12the screen. Ten months after the first
- 23:16known episode of the Therac 25, the
- 23:18failure leading to a massive overdose
- 23:20was finally reproduced. I haven't found
- 23:25information anywhere on how many people
- 23:27in total were treated on all the
- 23:28machines during that time. Heiger began
- 23:31measuring the dose emitted during the
- 23:33failure. The first result was 804 rads,
- 23:36but he quickly realized the beam was so
- 23:38intense that it was overloading the
- 23:40measuring equipment. He adjusted the
- 23:44sensors and received a result exceeding
- 23:464,000 rads. So, it isn't the
- 23:48manufacturer discovering what really
- 23:50happened, because the company still
- 23:51cannot replicate the error. It was
- 23:54Heager who taught them that sequences
- 23:56must be performed quickly. Only then
- 23:59were the AECL engineers also able to
- 24:01trigger the malfunction. Measurements
- 24:03conducted by the manufacturer showed
- 24:04that in the center of the field, the
- 24:06device could deliver as much as 25,000
- 24:08rads. The exact doses received by the
- 24:10patients remain unknown, because they
- 24:12varied wildly during subsequent
- 24:14simulations, so it's not entirely clear
- 24:16who got what. But it is known that all
- 24:19the victims received far too much.
- 24:22Chapter five. 8 seconds. You might ask,
- 24:34what was this actually about? Where did
- 24:36the problem lie? The Therac accelerates
- 24:38electrons in both modes, right? The
- 24:41catastrophe occurred when the Therac
- 24:43combined elements of one operation with
- 24:44the settings of another. After
- 24:49initially selecting X, the computer
- 24:51began preparing the device to produce
- 24:53X-rays, meaning that stronger beam.
- 24:57Setting the magnets took about 8
- 24:59seconds before everything rotated into
- 25:00place. If, within that time window, in
- 25:05the middle of those 8 seconds, the
- 25:07operator changed the settings to e, the
- 25:09target and filter would move out of the
- 25:10beam's path because they aren't
- 25:12supposed to be there, but the energy
- 25:14remains. So the accelerator behaves
- 25:20according to the original X setting,
- 25:21but the safety measures get out of its
- 25:23way. The Therac is, as it were, in two
- 25:28different states at once. A
- 25:30Schrödinger's Therac. The operator saw
- 25:32electron therapy on the screen. The
- 25:35rotating mechanism was also in the
- 25:37position intended for that type of
- 25:39treatment, but the accelerator was
- 25:41blasting the massive stream needed for
- 25:42X-rays. But energy isn't everything.
- 25:46The problem was also the concentration
- 25:48of the beam, which in these two
- 25:50settings was not scattered and, like an
- 25:52electron blade, was focused on a small
- 25:54area. It was cutting through the
- 25:56patient's body with electrons, causing
- 25:58terrible destruction. I don't know if I
- 26:00can put it that way. I think you
- 26:02understand for the sake of imagination.
- 26:05A huge number of electrons also
- 26:06overloaded the ionization chambers,
- 26:08which measured the dose. The sensors
- 26:11stopped working properly and informed
- 26:13the computer that the patient was
- 26:14receiving a fraction. That's why the
- 26:17Therac displayed "malfunction 54," but
- 26:19simultaneously suggested that there had
- 26:21been an underdose. So the operator had
- 26:24the option to press "Set" again,
- 26:28injuring the patient. According to the
- 26:31AECL engineer, the sound resembling
- 26:33frying eggs that he heard came
- 26:34precisely from those damaged and
- 26:36overloaded ionization chambers. And
- 26:40I'll be honest, when we first read that
- 26:42, I thought it might be about the
- 26:43denaturing of proteins in the body.
- 26:46Fortunately, no. So the whole problem
- 26:49was contained within eight seconds.
- 26:52Several procedures were running
- 26:53simultaneously and using the same
- 26:55information. The result depended on
- 26:57which one would be first. A race of
- 26:59sorts. That is exactly what is called a
- 27:01race condition. If the operator worked
- 27:03a bit slower, wasn't quite as efficient
- 27:05, the program would register the
- 27:07correction and correctly prepare
- 27:08everything for work. The error,
- 27:11therefore, had a very perverse
- 27:12characteristic. It was most easily
- 27:15triggered by the person who knew the
- 27:17device best and operated it the fastest
- 27:18. The second patient, Styler, died on
- 27:21May 1st, less than three weeks after
- 27:23the accident. The autopsy revealed
- 27:25acute radiation damage to the right
- 27:27temporal lobe and the brain stem. brain
- 27:30. Besides, the first patient's
- 27:31condition was also deteriorating. He
- 27:33lost control of his left arm. He
- 27:35suffered paralysis of both legs, one
- 27:37vocal cord, and his diaphragm. He died
- 27:39five months after the accident. Chapter
- 27:42six. Tape. Four days later, AECL sent
- 27:53Therac users instructions for a
- 27:55temporary fix to the problem. What you
- 27:57are about to hear now is not a joke. We
- 27:59have confirmed this through several
- 28:01sources. You had to take a screwdriver
- 28:04to the keyboard, remove the cursor-up
- 28:06key, and disable the contacts
- 28:08underneath with electrical tape. If the
- 28:12operator made a mistake while entering
- 28:14parameters, they had to reset the
- 28:16device and type everything in from the
- 28:17start. The primary protection against a
- 28:21fatal software bug thus became a
- 28:22ripped-out button and a piece of tape.
- 28:26The FDA deemed this reaction
- 28:27insufficient. Unsurprisingly, the
- 28:30letter to users did not explain the
- 28:32nature of the defect or what kind of
- 28:34danger it posed. On May 2, 1986, the
- 28:38FDA declared the Therac a defective
- 28:40device and ordered a corrective action
- 28:42plan. It was to include software
- 28:46changes, additional hardware safety
- 28:48measures, a redesign of the dose
- 28:50measurement system, and replacing
- 28:52cryptic error codes. With something
- 28:55clearer, something understandable. On
- 28:57January 17, 1987, another accident
- 28:59occurred at the Yakima Valley Memorial
- 29:02Hospital. The patient was to receive
- 29:05two small control exposures of four and
- 29:07three rads. A film was placed under his
- 29:10body, which was meant to record the
- 29:11radiation field. This allowed the staff
- 29:14to check if the device was aiming at
- 29:16the planned spot. After taking the
- 29:18images, the patient was supposed to
- 29:20receive another 79 rads of X-ray
- 29:22radiation. So that first shot was just
- 29:24for control. The operator delivers the
- 29:26first four rads. The Therac goes into a
- 29:28pause state. The field size changes
- 29:29automatically and it delivers another
- 29:31three rads. Then the operator enters
- 29:34the room to remove the film and
- 29:35position the patient. Using the manual
- 29:38controller, he rotated the internal
- 29:40mechanisms of the Therac to the aiming
- 29:41position. At that moment, a mirror was
- 29:44in the path of the beam. It reflected
- 29:46ordinary light, which showed the
- 29:48location of the future radiation site
- 29:50on the body. After finishing the setup,
- 29:53the operator commanded the machine to
- 29:54prepare for actual treatment. He forgot
- 29:58to remove the film from under the
- 29:59patient, which will be important in
- 30:01this story. On the screen, a message
- 30:02appeared informing that the beam was
- 30:04ready. The operator starts the device.
- 30:07The Therac begins operation, but the
- 30:08system shows neither the dose nor the
- 30:10rate of delivery. After five or six
- 30:13seconds, the machine stops working. The
- 30:15device has gone into a pause state. So
- 30:18the operator presses "proceed." The
- 30:21Therac activates the beam for the
- 30:22second time. The machine stops again.
- 30:25This time, a "flatness" message appears
- 30:27, indicating a problem with the uniform
- 30:29distribution of radiation. The operator
- 30:31heard the patient's voice over the
- 30:33intercom, but didn't understand it. He
- 30:35entered the room. The man said he felt
- 30:37a strong burning sensation in his chest
- 30:39. The screen indicated that the patient
- 30:41had received a total of seven rads.
- 30:43Exactly the sum of those earlier
- 30:44control exposures. The system did not
- 30:48record the radiation emitted later.
- 30:51That same day, burns covering the
- 30:53entire treatment area appeared on the
- 30:54patient's skin. After four days, the
- 30:57redness had arranged itself into
- 30:59stripes. An identical pattern had
- 31:00already been observed a year earlier.
- 31:03At the time, if you recall, the cause
- 31:05was labeled as unknown. AECL begins an
- 31:08investigation. However, it was
- 31:11suspected that the beam might have been
- 31:13fired while the turntable was still in
- 31:15the setup position, and a forgotten
- 31:17film came in handy. A hospital
- 31:21physicist placed another film under the
- 31:23device, left the turntable in the setup
- 31:25position, and ran the Therac with X-ray
- 31:27parameters. He obtained a mark
- 31:31resembling a striped pattern. After a
- 31:35week of checking the equipment, AECL
- 31:37admitted that the machine's malfunction
- 31:39was not caused by a mechanical failure.
- 31:43They then began analyzing the software
- 31:45and found more errors. The program
- 31:48repeatedly checked whether the device's
- 31:50components were in the correct position
- 31:52. It used a variable for this, called
- 31:54class 3. A non-zero value meant the
- 31:57mechanism's position needed to be
- 31:59checked. Zero allowed the check to be
- 32:01bypassed, assuming everything was fine.
- 32:03Forgive me for saying this, but you
- 32:05will soon understand what it's about.
- 32:07The programmer wanted the variable to
- 32:09remain non-zero during treatment
- 32:11preparation. Instead of assigning it a
- 32:14constant value, like one, the program
- 32:17incremented it during every test: 1, 2,
- 32:203, 4. The variable occupied 1 byte of
- 32:24memory. It could therefore store
- 32:27numbers between 0 and 255. Once the
- 32:31device had been running long enough,
- 32:34upon reaching 255, further incrementing
- 32:39reset the counter to zero. Every 256th
- 32:47pass through this part of the program,
- 32:49the variable briefly displayed zero.
- 32:52The system skipped the position check,
- 32:54and during the accident, the operator
- 32:56pressed "set." At exactly that moment,
- 32:59the Therac allowed the accelerator to
- 33:01start, even though the turntable was in
- 33:03the setup position. Thus, the mirror
- 33:06remained in the path of the beam. The
- 33:08target, the filter, and the electron
- 33:10scattering system were missing. The
- 33:12ionization chamber that measures the
- 33:14dose was also not in that position. The
- 33:16designers, after all, did not
- 33:18anticipate that one could irradiate
- 33:19someone then. The Therac launched a
- 33:22powerful 25 megaelectron-volt stream.
- 33:24The beam was scattered and deflected by
- 33:26the steel mirror, then headed straight
- 33:28for the patient. Later measurements
- 33:30indicated that a single activation
- 33:32could deliver 4,000 to 5,000 rads. The
- 33:36operator pressed "P" and fired the beam
- 33:38again. The patient could have received
- 33:40a total of 8,000 to 10,000. Let me
- 33:43remind you, a total of 86 had been
- 33:45planned. The patient died in '87 due to
- 33:48complications. He had previously
- 33:51suffered from a terminal form of cancer
- 33:52, but his loved ones claimed the
- 33:54failure shortened his life and
- 33:56subjected him to horrific pain until
- 33:57the very end. The accident forced
- 34:00personnel to investigate a patient
- 34:02treated in Yakima in December '85. The
- 34:06woman had chronic ulcers and tissue
- 34:08necrosis, suffered from pain, and
- 34:09underwent surgeries and skin grafts.
- 34:12She survived, but was left with scars
- 34:14and disability. What is particularly
- 34:18bitter is that the hardware and
- 34:19software changes planned after the
- 34:21accidents could have prevented this
- 34:22event. Nine months later, they had
- 34:27still not been installed in the machine
- 34:29working in Yakima. The first known bug
- 34:32occurred within 8 seconds. The next one
- 34:34waited until the counter reached 256
- 34:38and went to zero. Chapter seven, class
- 34:40one. After another accident, the FDA
- 34:51decided it had enough. Continuing to
- 34:54use the Theracs while fixes were still
- 34:56being worked on by engineers was a
- 34:58dangerous makeshift solution. On
- 35:00February 6, 1987, a representative from
- 35:02the American agency contacted Canadian
- 35:05authorities. Both sides agreed that the
- 35:08machines had to be shut down until
- 35:09permanent safeguards were in place.
- 35:14Four days later, the FDA demanded that
- 35:17AECL inform all centers that the Therac
- 35:19-25 could not operate or be used for
- 35:21therapy. Exceptions were allowed only
- 35:27in cases where the need to treat a
- 35:28specific patient outweighed the risks
- 35:30of using the machine. And no one was
- 35:34under the illusion anymore that all the
- 35:36bugs had been found. In March,
- 35:38representatives from AECL, all clinics
- 35:41using the Theracs, and American and
- 35:43Canadian officials and medical
- 35:44physicists met at one table. They
- 35:47analyzed six accidents and prepared a
- 35:49list of changes. Users demanded, among
- 35:52other things, an independent software
- 35:54assessment and a permanent record of
- 35:56the device’s operation to allow for
- 35:58the reconstruction of its internal
- 35:59state after a potential accident. The
- 36:03AECL representative admitted that tests
- 36:05of the fixes had been performed, but
- 36:06they were not documented. He also
- 36:09stated that an independent software
- 36:11assessment might prove impossible. The
- 36:14company had no intention of releasing
- 36:15the source code. On May 26, the FDA
- 36:19conditionally approved the fourth
- 36:21version of the corrective plan and
- 36:23classified the case as Class I. This
- 36:25was the highest class of product recall
- 36:27, used when there is a danger that the
- 36:29use of a device could cause injury. In
- 36:33June, a new safety analysis showed that
- 36:36the plan still did not cover
- 36:37potentially dangerous areas such as
- 36:39beam scanning, energy selection, beam
- 36:41shutdown, and control calibration. A
- 36:44fifth version of the corrective plan
- 36:46was needed. On July 21, 1987, AECL
- 36:49presented its final corrective plan.
- 36:53From then on, if an error was detected,
- 36:55the machine would completely abort the
- 36:56treatment without the possibility of
- 36:58resuming with the P key. Independent
- 37:01hardware safeguards were also added to
- 37:03stop the dangerous beam and prevent it
- 37:05from firing if the machine was
- 37:07misconfigured. Enigmatic fault codes
- 37:09were replaced with understandable
- 37:11messages, and the function of editing
- 37:13keys was restricted. The set and reset
- 37:15commands were changed. Additional
- 37:17software interlocks were added, and
- 37:19instructions were improved. The full
- 37:21implementation of the safeguards, along
- 37:23with upgrades and safety analysis, took
- 37:25another year. The final report was
- 37:27released on November 3, 1988. Two years
- 37:30and seven months had passed since the
- 37:32Tyler accident, and three years and
- 37:34five months since the overdose in
- 37:36Kennestone. All machines were
- 37:38re-approved for treating patients.
- 37:41After the additional safeguards were
- 37:43introduced, no more accidents involving
- 37:44the machines were ever recorded. It is
- 37:47unknown how much longer they remained
- 37:49in operation. Let’s summarize.
- 38:01Stories like the Therac-25 fuel
- 38:03distrust of medicine, and sometimes it
- 38:05is hard to be surprised by that.
- 38:08Patients entrusted their lives to
- 38:09specialists and devices that were
- 38:11supposed to heal them. Meanwhile, the
- 38:13manufacturer spent months claiming that
- 38:15a massive overdose was impossible. It
- 38:18is easy then to think: if doctors,
- 38:20engineers, and institutions make
- 38:22mistakes, perhaps it is safer to look
- 38:24for help outside the system, from
- 38:26people offering miracle therapies,
- 38:28detoxes, healing energy, or vibrations.
- 38:31However, an unproven method does not
- 38:33become effective just because modern
- 38:35medicine can be fallible. The Terak 25
- 38:38was rebuilt and is back in use for
- 38:39treatment. Locks were added, programs
- 38:42were improved, and so on. There have
- 38:43been no further accidents reported. Of
- 38:46course, for earlier victims, it's too
- 38:49late and changes nothing, but let's
- 38:51remember that the rise in life
- 38:53expectancy is owed to thousands of such
- 38:55steps, errors, problems, and later
- 38:57corrections, clean water, vaccines,
- 38:59antibiotics, and safe surgeries. You
- 39:05know, after such accidents, there’s
- 39:07usually a revised manual, an extra
- 39:09sensor, a duty to report failures,
- 39:11millions of minor changes that add up
- 39:13to extending human life. Medicine must
- 39:18earn trust by disclosing errors,
- 39:20investigating them, and turning them
- 39:21into safeguards that didn't exist
- 39:23before. This is the only way for people
- 39:28to trust medicine on a mass scale.
- 39:31Thank you for the highest proof of
- 39:33appreciation, which is staying with us
- 39:34until the end. We are extremely pleased
- 39:36that there are more and more of you.
- 39:39Have a good day, stay healthy, and see
- 39:41you next time.
About this transcript
This page contains the full transcript of Najgroźniejszy błąd w historii radioterapii by OdwróconaLekcja, generated from the public captions YouTube serves with the video. The transcript has 6,105 words across 1,001 segments, with the original timestamps preserved so you can click any line to jump to that moment in the embedded player.
What you can do with it
Use the transcript to take notes, quote the speaker, build a study guide, generate a summary with ChatGPT or Claude via the YouTube Summary tool, or export it as a timed subtitle file with YouTube to SRT. You can also re-open it in the transcriber to translate the transcript into 100+ languages.