- 0:00can you imagine working in a critical
- 0:02care environment without pulse oximetry
- 0:05oximetry was invented in 1935 but it
- 0:09wasn't until the pulse oximeter was
- 0:11developed in 1972 and released
- 0:14commercially in 1980 that this became a
- 0:16universal critical care monitoring
- 0:19modality my name is Ken Hoffman I'm an
- 0:22intensive care specialist at the Alfred
- 0:25hospital in Melbourne in this video
- 0:27we're going to review how this
- 0:29incredible piece of technology works
- 0:31this video is made for those sitting the
- 0:34Australian and New Zealand College of
- 0:36intensive care medicine first part
- 0:38examination to demonstrate the expected
- 0:40level of knowledge as with all
- 0:43measurement and monitoring devices there
- 0:45are four components to the system one a
- 0:48biological variable - a sensor 3 an
- 0:54integrator and 4 an output
- 0:59firstly the biological variable that we
- 1:02are measuring is the percentage
- 1:04saturation of hemoglobin molecules in
- 1:07the bloodstream
- 1:09to calculate this we rely on the
- 1:11principle that oxygenated and
- 1:13deoxygenated hemoglobin have different
- 1:17light absorption characteristics
- 1:18depending on the light wavelength we can
- 1:23see this visually because oxygenated
- 1:25hemoglobin is a bright red color and
- 1:28deoxygenated hemoglobin is a deep dark
- 1:31red
- 1:33the next component is the sensor system
- 1:36this consists of a pair of
- 1:39light-emitting diodes which emit
- 1:41monochromatic red light at a wavelength
- 1:44of 660 nanometers and monochromatic
- 1:47infrared light which we cannot see
- 1:50because it's outside the visible
- 1:52spectrum at a wavelength of 914
- 1:55nanometers
- 1:56the reason these wavelengths were
- 1:59selected is because at these wavelengths
- 2:01oxygenated hemoglobin and deoxygenated
- 2:04hemoglobin have very different absorb
- 2:07ative properties this allows us to
- 2:10calculate the proportion of oxygenated
- 2:13hemoglobin and deoxygenated hemoglobin
- 2:15that is present we then have a pair of
- 2:18photo diodes or light sensors which
- 2:22detect how much light of each wavelength
- 2:25has passed through the tissues and
- 2:27converts this to an electrical signal
- 2:30these sensors are surrounded by an
- 2:33opaque case to minimize the ambient
- 2:36light signal next we have the integrator
- 2:40or computer which processes the
- 2:42electrical signal this is where pulse
- 2:45oximetry gets very clever
- 2:50the signal we need to review to physics
- 2:52laws beers law and Lambert's law these
- 2:56are often combined and termed beer
- 2:59Lambert's law beers law states that the
- 3:03absorption or attenuation of life as it
- 3:07passes through a substance is
- 3:09proportional to the concentration of
- 3:11that substance this is how the pulse
- 3:14oximeter calculates the concentration or
- 3:17relative proportions of oxygenated
- 3:21hemoglobin and deoxygenated hemoglobin
- 3:25the second law is Lambert's law this
- 3:28states that the absorption or
- 3:30attenuation of light as it passes
- 3:32through a substance is proportional to
- 3:35the distance the light has to travel
- 3:37this is how the pulse oximeter can
- 3:41isolate the pulsatile component of the
- 3:44signal as blood moves into the tissues
- 3:47with each cardiac contraction there is a
- 3:50tiny but measurable increase in the
- 3:53distance between the light emitting
- 3:55diodes and the photodiode detectors we
- 3:59can then isolate the pulsatile component
- 4:01of the signal which reflects arterial
- 4:04blood entering the tissues
- 4:07we do this by assessing that aec or
- 4:10pulsatile component of the signal
- 4:12compared to the ratio of DC or
- 4:16non-pulsatile component of the signal we
- 4:20do this for both six hundred and sixty
- 4:22nanometer and 940 nanometer light
- 4:25wavelengths we then combine these ratios
- 4:28for both six hundred and sixty nanometer
- 4:30and 940 nanometer light into another
- 4:34ratio termed the R value once the
- 4:38computer has calculated the R value and
- 4:41then converts the R value into an oxygen
- 4:44saturation level obtained from
- 4:46experimental data where healthy
- 4:48volunteers braved hypoxic gas mixtures
- 4:51this is the reason why pulse oximeters
- 4:54are calibrated down to a saturation
- 4:57level of around 75 percent beyond this
- 5:01point the curve is extrapolated as it's
- 5:03very hard to get ethics approval or
- 5:05volunteers willing to become this
- 5:07hypoxic
- 5:09at this point we should mention a couple
- 5:11of clever things
- 5:12the pulse oximeter does to clean up the
- 5:14signal
- 5:16the first is to remove interference from
- 5:19ambient light sources
- 5:21it does this bicycling through which
- 5:24light emitting diode is actually
- 5:26emitting light 660 nanometer and then
- 5:30940 nanometer and then having both of
- 5:34the light sources turned off we can't
- 5:37actually see this as the light sources
- 5:39are flickering at up to 900 Hertz during
- 5:44the time that both light emitting diodes
- 5:45are off the computer calculates the
- 5:48effect of ambient light on the
- 5:50photodiode detector it could then remove
- 5:53ambient light from the analysis of the
- 5:56signal the second is that after ambient
- 5:59light and non-pulsatile signal is
- 6:01removed
- 6:02the pulse oximeter amplifies the
- 6:04positive signal and takes an average
- 6:06over a few seconds before calculating
- 6:09the saturation this is to try and
- 6:12identify and compensate for any movement
- 6:15artifact that has occurred
- 6:18the final component of the system is the
- 6:20output this includes a numerical display
- 6:23usually with the saturations and the
- 6:26heartrate and on higher quality monitors
- 6:29a graph called a plasma graph which
- 6:32means a graph of the changing volume
- 6:35this allows an assessment of the quality
- 6:38of the signal
- 6:39so to recap our pulse oximeter consists
- 6:43of four components one a biological
- 6:47variable which is the percentage
- 6:50saturation of hemoglobin 2 a sensor
- 6:54which is assessing the transmission of
- 6:57660 nanometer and 940 nanometer
- 7:01wavelengths of light through the tissues
- 7:043 an integrator which is analyzing the
- 7:09signal from the sensor and separating
- 7:11out the pulsatile component of the
- 7:13signal and 4 an output which is a
- 7:18numerical display of the saturations and
- 7:20heart rate and a graphical display of
- 7:22the quality of the signal
- 7:26thank you for listening if this was
- 7:28useful please hit the like button if
- 7:30you'd like to see more videos like this
- 7:33please subscribe to our Channel
About this transcript
This page contains the full transcript of How does Pulse Oximetry work? by The Alfred Intensive Care Academic Centre, generated from the public captions YouTube serves with the video. The transcript has 945 words across 162 segments, with the original timestamps preserved so you can click any line to jump to that moment in the embedded player.
What you can do with it
Use the transcript to take notes, quote the speaker, build a study guide, generate a summary with ChatGPT or Claude via the YouTube Summary tool, or export it as a timed subtitle file with YouTube to SRT. You can also re-open it in the transcriber to choose a caption track available on the video.
Free YouTube transcript tool
YouTube2Text is a free YouTube transcript generator — no signup, no daily limit. Paste any YouTube link and get the full transcript instantly, with timestamps, click-to-jump, caption language selection, AI prompts for ChatGPT, Claude, and Gemini, and exports to TXT, SRT, VTT, or Markdown.