Good Microbiological Practices and Procedures (GMPP) 5: autoclaves — Transcript
Full transcript
- 0:05of all the methods surveillance to
- 0:08decontaminate laboratory waste
- 0:10autoclaving is one of the easiest and
- 0:13most effective autoclaving combines heat
- 0:16steam and high pressure to kill
- 0:19biological agents it's important to note
- 0:23that some biological agents are more
- 0:25robust than others for example bacterial
- 0:29spores and prions pretreatment with
- 0:32disinfectant increased temperature
- 0:34pressure and contact time may be
- 0:37required to neutralize such agents
- 0:40different combinations of temperature
- 0:43and contact time are used depending on
- 0:46the types of materials being sterilized
- 0:48a fundamental requirement is the need
- 0:51for the heat to penetrate throughout a
- 0:53load dense loads or ones with high
- 0:57organic content are hardest to sterilize
- 1:00because they resist the penetration of
- 1:03the steam and heat glassware is the
- 1:07easiest because there is a lower load
- 1:10density and it traps heat within the
- 1:12contents steam is generated at the
- 1:16beginning of a cycle delivered either by
- 1:19boiling water in a reservoir or
- 1:21injecting piped steam depending on the
- 1:24type of autoclave and facilities
- 1:26available the autoclave cycle consists
- 1:31of three phases in the heat up phase
- 1:35temperature and pressure are raised to
- 1:38preset values some autoclaves create a
- 1:42partial vacuum during this phase to help
- 1:45steam penetrate the load
- 1:47during the dwell time or holding time
- 1:50the autoclave maintains a constant
- 1:53temperature and pressure to ensure
- 1:55sterilization
- 1:58during the cooling down phase heat and
- 2:01pressure are allowed to fall to ambient
- 2:03levels in some autoclaves there may be
- 2:06an assisted cooling function to reduce
- 2:08the overall cycle time
- 2:12some typical autoclave parameters are
- 2:16115 Celsius for 30 minutes 121 Celsius
- 2:21for 15 minutes or a hundred and 15
- 2:24Celsius for 60 minutes these are all
- 2:27validated to achieve sterilization of
- 2:30most biological agents assuming
- 2:32sufficient penetration but spores and
- 2:35prions may require higher temperatures
- 2:38and all pretreatment the dwell time may
- 2:41be increased to ensure sufficient
- 2:43penetration in denser loads or generally
- 2:46as an additional precautionary measure
- 2:50in modern autoclaves there is usually a
- 2:53safety thermal lock which prevents the
- 2:56lid or door being opened at scalding
- 2:58temperature if the autoclave has a
- 3:02printout that displays the temperature
- 3:04it may display the sterilization cycle
- 3:07as a graph or as the temperatures
- 3:10actually achieved these may fluctuate
- 3:13slightly above and below the set
- 3:15temperature but this is normal as the
- 3:18heater pulses so long as it achieves the
- 3:20average over the time period when the
- 3:23autoclave is serviced this will be
- 3:25checked and validated it's important to
- 3:29note that when autoclaving bottles
- 3:31either full or empty it's advisable to
- 3:34loosen the caps to ensure that pressure
- 3:37differentials don't cause them to break
- 3:40waste to be autoclaved must be packaged
- 3:43such that steam can penetrate the entire
- 3:46load plastic autoclave bags must be
- 3:49loosely taped ideally with indicator
- 3:52tape and not tightly bound these should
- 3:55then be placed in autoclave herbal boxes
- 3:58to contain their contents as the plastic
- 4:01bags may melt and leak some packaging
- 4:04such as paper autoclave bags often used
- 4:07for sterilizing small items like steel
- 4:10coupons and spatulas is designed so that
- 4:13the steam will permeate through the
- 4:15packaging which remains completely
- 4:17sealed appropriate PPE such as a
- 4:24laboratory coat and gloves and eye
- 4:27protection for handling liquids should
- 4:29be worn when handling waste and when
- 4:32transporting waste to the autoclave
- 4:35items should not be stacked on top of
- 4:37one another as this prevents steam
- 4:40penetration unless there is shelving or
- 4:43baskets to create space between the
- 4:45items
- 4:48the autoclave cycle should be verified
- 4:51before first use and periodically
- 4:54thereafter to make sure the cycle
- 4:57parameters are effectively killing the
- 4:59microorganisms handled in your
- 5:01laboratory some autoclaves can be fitted
- 5:05with load sensing probes which can be
- 5:08inserted into the center of a dense load
- 5:10to monitor temperature and verify that
- 5:13sterilization temperature has been
- 5:15achieved chemical indicators such as
- 5:19autoclave tape use heat sensitive ink
- 5:22that changes color when it reaches a set
- 5:25temperature this is a useful indicator
- 5:27but it's only checking the temperatures
- 5:30on the outside surfaces not the contents
- 5:32and it doesn't tell you if the required
- 5:35pressure and dwell time were achieved
- 5:38another method of verification is to put
- 5:42a broth culture of an appropriate
- 5:43biological agent through the autoclave
- 5:46cycle if subsequent attempts to grow the
- 5:49biological agent from the autoclave
- 5:51broth fail to produce growth then the
- 5:54autoclave cycle has been shown to be
- 5:56effective biological indicators such as
- 6:02spore strips or discs may also be used
- 6:04to provide verification there Oh bust
- 6:08nature means that if they are killed by
- 6:10the autoclave cycle then most other
- 6:12biological agents will be killed by the
- 6:15same cycle spore strips or discs are
- 6:19generally placed at various locations
- 6:21within the autoclave to ensure that all
- 6:24areas reach the required parameters they
- 6:27can also be placed within a typical
- 6:29autoclave load although it can prove
- 6:32difficult to extract them from the load
- 6:34when the autoclave cycle is complete the
- 6:39recovered strips or discs are placed in
- 6:41liquid broth and incubated
- 6:45no growth after a set time period such
- 6:49as 48 hours would mean the autoclave
- 6:51cycle is effective autoclaves must be
- 6:56well-maintained
- 6:57to make sure they perform as required
- 7:00this increases productivity and reduces
- 7:03downtime routine in use maintenance
- 7:07includes making sure any water
- 7:10reservoirs are kept clean and filled
- 7:12with water and are regularly checked for
- 7:15signs of wear and tear such as corrosion
- 7:18the temperature profile of each
- 7:20autoclave load must be checked to make
- 7:23sure that the sterilization conditions
- 7:25have been achieved this is important as
- 7:28if the waste has not been sterilized
- 7:31other people handling the waste post
- 7:34autoclaving are at risk of exposure and
- 7:37subsequent potential infection
- 7:42autoclaves should be serviced at least
- 7:44annually and tested by a specialist
- 7:47contractor who can make sure the set
- 7:49temperature and pressure cycles are
- 7:51accurate and check the overall condition
- 7:54of the autoclave this is important
- 7:57because of the high pressures at which
- 7:59they operate because of these high
- 8:02pressures in some countries it is also a
- 8:04statutory requirement for autoclaves to
- 8:07be tested by an independent pressure
- 8:10systems Assessor if there is a power
- 8:14failure or if the autoclave breaks down
- 8:17during a sterilization cycle the
- 8:19autoclave should not be opened straight
- 8:21away instead attempts should be made to
- 8:25fix the autoclave and repeat the cycle
- 8:30only if it's not possible to restart
- 8:32should the autoclave be open and the
- 8:35incompletely autoclaved waste removed
- 8:39the waste removed from the autoclave may
- 8:43still contain live biological agents so
- 8:46you will need to wear the same PPE that
- 8:48you used when loading the autoclave this
- 8:52waste must be suitably contained for
- 8:54storage and should be labeled to
- 8:57indicate that it is an autoclave waste
- 9:01if the waste needs to be sent elsewhere
- 9:04for sterilization it will need to be
- 9:07suitably packaged for transportation in
- 9:09accordance with the UN model regulations
- 9:14once the waste has been removed if
- 9:17access to the autoclave chamber is
- 9:19required for maintenance or repair the
- 9:21inside of it should be wiped down with
- 9:24an effective disinfectant in summary
- 9:29good practices when using autoclaves
- 9:32include wear appropriate PPE when
- 9:36loading and unloading verify cycle
- 9:40parameters are being achieved by
- 9:42checking temperature profiles and use
- 9:45load sensing probes where possible
- 9:48ensure steam penetration by taping bags
- 9:52loosely and not tightly packing or over
- 9:55filling the autoclave loosened lids of
- 9:59bottles before starting the autoclave
- 10:02use autoclave boxes or tins to contain
- 10:06any melted items or liquids use chemical
- 10:11and biological indicators to check that
- 10:14cycle parameters are effective for the
- 10:16biological agents you are handling have
- 10:20a regular maintenance program in place
- 10:28[Music]
- 10:32you
About this transcript
This page contains the full transcript of Good Microbiological Practices and Procedures (GMPP) 5: autoclaves by World Health Organization (WHO), generated from the public captions YouTube serves with the video. The transcript has 1,264 words across 210 segments, with the original timestamps preserved so you can click any line to jump to that moment in the embedded player.
What you can do with it
Use the transcript to take notes, quote the speaker, build a study guide, generate a summary with ChatGPT or Claude via the YouTube Summary tool, or export it as a timed subtitle file with YouTube to SRT. You can also re-open it in the transcriber to translate the transcript into 100+ languages.
Free YouTube transcript tool
YouTube2Text is a free YouTube transcript generator — no signup, no daily limit. Paste any YouTube link and get the full transcript instantly, with timestamps, click-to-jump, translation to 100+ languages, AI prompts for ChatGPT, Claude, and Gemini, and exports to TXT, SRT, VTT, or Markdown.