Good Microbiological Practices and Procedures (GMPP) 5: autoclaves — Transcript

by World Health Organization (WHO) · 1,264 words · 210 segments · language en · Watch on YouTube

Full transcript

  1. 0:05of all the methods surveillance to
  2. 0:08decontaminate laboratory waste
  3. 0:10autoclaving is one of the easiest and
  4. 0:13most effective autoclaving combines heat
  5. 0:16steam and high pressure to kill
  6. 0:19biological agents it's important to note
  7. 0:23that some biological agents are more
  8. 0:25robust than others for example bacterial
  9. 0:29spores and prions pretreatment with
  10. 0:32disinfectant increased temperature
  11. 0:34pressure and contact time may be
  12. 0:37required to neutralize such agents
  13. 0:40different combinations of temperature
  14. 0:43and contact time are used depending on
  15. 0:46the types of materials being sterilized
  16. 0:48a fundamental requirement is the need
  17. 0:51for the heat to penetrate throughout a
  18. 0:53load dense loads or ones with high
  19. 0:57organic content are hardest to sterilize
  20. 1:00because they resist the penetration of
  21. 1:03the steam and heat glassware is the
  22. 1:07easiest because there is a lower load
  23. 1:10density and it traps heat within the
  24. 1:12contents steam is generated at the
  25. 1:16beginning of a cycle delivered either by
  26. 1:19boiling water in a reservoir or
  27. 1:21injecting piped steam depending on the
  28. 1:24type of autoclave and facilities
  29. 1:26available the autoclave cycle consists
  30. 1:31of three phases in the heat up phase
  31. 1:35temperature and pressure are raised to
  32. 1:38preset values some autoclaves create a
  33. 1:42partial vacuum during this phase to help
  34. 1:45steam penetrate the load
  35. 1:47during the dwell time or holding time
  36. 1:50the autoclave maintains a constant
  37. 1:53temperature and pressure to ensure
  38. 1:55sterilization
  39. 1:58during the cooling down phase heat and
  40. 2:01pressure are allowed to fall to ambient
  41. 2:03levels in some autoclaves there may be
  42. 2:06an assisted cooling function to reduce
  43. 2:08the overall cycle time
  44. 2:12some typical autoclave parameters are
  45. 2:16115 Celsius for 30 minutes 121 Celsius
  46. 2:21for 15 minutes or a hundred and 15
  47. 2:24Celsius for 60 minutes these are all
  48. 2:27validated to achieve sterilization of
  49. 2:30most biological agents assuming
  50. 2:32sufficient penetration but spores and
  51. 2:35prions may require higher temperatures
  52. 2:38and all pretreatment the dwell time may
  53. 2:41be increased to ensure sufficient
  54. 2:43penetration in denser loads or generally
  55. 2:46as an additional precautionary measure
  56. 2:50in modern autoclaves there is usually a
  57. 2:53safety thermal lock which prevents the
  58. 2:56lid or door being opened at scalding
  59. 2:58temperature if the autoclave has a
  60. 3:02printout that displays the temperature
  61. 3:04it may display the sterilization cycle
  62. 3:07as a graph or as the temperatures
  63. 3:10actually achieved these may fluctuate
  64. 3:13slightly above and below the set
  65. 3:15temperature but this is normal as the
  66. 3:18heater pulses so long as it achieves the
  67. 3:20average over the time period when the
  68. 3:23autoclave is serviced this will be
  69. 3:25checked and validated it's important to
  70. 3:29note that when autoclaving bottles
  71. 3:31either full or empty it's advisable to
  72. 3:34loosen the caps to ensure that pressure
  73. 3:37differentials don't cause them to break
  74. 3:40waste to be autoclaved must be packaged
  75. 3:43such that steam can penetrate the entire
  76. 3:46load plastic autoclave bags must be
  77. 3:49loosely taped ideally with indicator
  78. 3:52tape and not tightly bound these should
  79. 3:55then be placed in autoclave herbal boxes
  80. 3:58to contain their contents as the plastic
  81. 4:01bags may melt and leak some packaging
  82. 4:04such as paper autoclave bags often used
  83. 4:07for sterilizing small items like steel
  84. 4:10coupons and spatulas is designed so that
  85. 4:13the steam will permeate through the
  86. 4:15packaging which remains completely
  87. 4:17sealed appropriate PPE such as a
  88. 4:24laboratory coat and gloves and eye
  89. 4:27protection for handling liquids should
  90. 4:29be worn when handling waste and when
  91. 4:32transporting waste to the autoclave
  92. 4:35items should not be stacked on top of
  93. 4:37one another as this prevents steam
  94. 4:40penetration unless there is shelving or
  95. 4:43baskets to create space between the
  96. 4:45items
  97. 4:48the autoclave cycle should be verified
  98. 4:51before first use and periodically
  99. 4:54thereafter to make sure the cycle
  100. 4:57parameters are effectively killing the
  101. 4:59microorganisms handled in your
  102. 5:01laboratory some autoclaves can be fitted
  103. 5:05with load sensing probes which can be
  104. 5:08inserted into the center of a dense load
  105. 5:10to monitor temperature and verify that
  106. 5:13sterilization temperature has been
  107. 5:15achieved chemical indicators such as
  108. 5:19autoclave tape use heat sensitive ink
  109. 5:22that changes color when it reaches a set
  110. 5:25temperature this is a useful indicator
  111. 5:27but it's only checking the temperatures
  112. 5:30on the outside surfaces not the contents
  113. 5:32and it doesn't tell you if the required
  114. 5:35pressure and dwell time were achieved
  115. 5:38another method of verification is to put
  116. 5:42a broth culture of an appropriate
  117. 5:43biological agent through the autoclave
  118. 5:46cycle if subsequent attempts to grow the
  119. 5:49biological agent from the autoclave
  120. 5:51broth fail to produce growth then the
  121. 5:54autoclave cycle has been shown to be
  122. 5:56effective biological indicators such as
  123. 6:02spore strips or discs may also be used
  124. 6:04to provide verification there Oh bust
  125. 6:08nature means that if they are killed by
  126. 6:10the autoclave cycle then most other
  127. 6:12biological agents will be killed by the
  128. 6:15same cycle spore strips or discs are
  129. 6:19generally placed at various locations
  130. 6:21within the autoclave to ensure that all
  131. 6:24areas reach the required parameters they
  132. 6:27can also be placed within a typical
  133. 6:29autoclave load although it can prove
  134. 6:32difficult to extract them from the load
  135. 6:34when the autoclave cycle is complete the
  136. 6:39recovered strips or discs are placed in
  137. 6:41liquid broth and incubated
  138. 6:45no growth after a set time period such
  139. 6:49as 48 hours would mean the autoclave
  140. 6:51cycle is effective autoclaves must be
  141. 6:56well-maintained
  142. 6:57to make sure they perform as required
  143. 7:00this increases productivity and reduces
  144. 7:03downtime routine in use maintenance
  145. 7:07includes making sure any water
  146. 7:10reservoirs are kept clean and filled
  147. 7:12with water and are regularly checked for
  148. 7:15signs of wear and tear such as corrosion
  149. 7:18the temperature profile of each
  150. 7:20autoclave load must be checked to make
  151. 7:23sure that the sterilization conditions
  152. 7:25have been achieved this is important as
  153. 7:28if the waste has not been sterilized
  154. 7:31other people handling the waste post
  155. 7:34autoclaving are at risk of exposure and
  156. 7:37subsequent potential infection
  157. 7:42autoclaves should be serviced at least
  158. 7:44annually and tested by a specialist
  159. 7:47contractor who can make sure the set
  160. 7:49temperature and pressure cycles are
  161. 7:51accurate and check the overall condition
  162. 7:54of the autoclave this is important
  163. 7:57because of the high pressures at which
  164. 7:59they operate because of these high
  165. 8:02pressures in some countries it is also a
  166. 8:04statutory requirement for autoclaves to
  167. 8:07be tested by an independent pressure
  168. 8:10systems Assessor if there is a power
  169. 8:14failure or if the autoclave breaks down
  170. 8:17during a sterilization cycle the
  171. 8:19autoclave should not be opened straight
  172. 8:21away instead attempts should be made to
  173. 8:25fix the autoclave and repeat the cycle
  174. 8:30only if it's not possible to restart
  175. 8:32should the autoclave be open and the
  176. 8:35incompletely autoclaved waste removed
  177. 8:39the waste removed from the autoclave may
  178. 8:43still contain live biological agents so
  179. 8:46you will need to wear the same PPE that
  180. 8:48you used when loading the autoclave this
  181. 8:52waste must be suitably contained for
  182. 8:54storage and should be labeled to
  183. 8:57indicate that it is an autoclave waste
  184. 9:01if the waste needs to be sent elsewhere
  185. 9:04for sterilization it will need to be
  186. 9:07suitably packaged for transportation in
  187. 9:09accordance with the UN model regulations
  188. 9:14once the waste has been removed if
  189. 9:17access to the autoclave chamber is
  190. 9:19required for maintenance or repair the
  191. 9:21inside of it should be wiped down with
  192. 9:24an effective disinfectant in summary
  193. 9:29good practices when using autoclaves
  194. 9:32include wear appropriate PPE when
  195. 9:36loading and unloading verify cycle
  196. 9:40parameters are being achieved by
  197. 9:42checking temperature profiles and use
  198. 9:45load sensing probes where possible
  199. 9:48ensure steam penetration by taping bags
  200. 9:52loosely and not tightly packing or over
  201. 9:55filling the autoclave loosened lids of
  202. 9:59bottles before starting the autoclave
  203. 10:02use autoclave boxes or tins to contain
  204. 10:06any melted items or liquids use chemical
  205. 10:11and biological indicators to check that
  206. 10:14cycle parameters are effective for the
  207. 10:16biological agents you are handling have
  208. 10:20a regular maintenance program in place
  209. 10:28[Music]
  210. 10:32you

About this transcript

This page contains the full transcript of Good Microbiological Practices and Procedures (GMPP) 5: autoclaves by World Health Organization (WHO), generated from the public captions YouTube serves with the video. The transcript has 1,264 words across 210 segments, with the original timestamps preserved so you can click any line to jump to that moment in the embedded player.

What you can do with it

Use the transcript to take notes, quote the speaker, build a study guide, generate a summary with ChatGPT or Claude via the YouTube Summary tool, or export it as a timed subtitle file with YouTube to SRT. You can also re-open it in the transcriber to translate the transcript into 100+ languages.

Free YouTube transcript tool

YouTube2Text is a free YouTube transcript generator — no signup, no daily limit. Paste any YouTube link and get the full transcript instantly, with timestamps, click-to-jump, translation to 100+ languages, AI prompts for ChatGPT, Claude, and Gemini, and exports to TXT, SRT, VTT, or Markdown.