YouTube2Text

Biological safety cabinet (BSC) 3: Best practices for safe usage — Transcript

by World Health Organization (WHO) · 651 words · 105 segments · language en · Watch on YouTube

Full transcript

  1. 0:03[Music]
  2. 0:05it is always important to follow best
  3. 0:08practices while working in a biosafety
  4. 0:10cabinet also known as a BSC these
  5. 0:14practices will help maintain adequate
  6. 0:16airflow within the cabinet to help
  7. 0:19prevent exposures and release of
  8. 0:20pathogens they should also be
  9. 0:22incorporated into your facility's
  10. 0:24standard operating procedures now let's
  11. 0:28take a look at some of these recommended
  12. 0:29practices work in a BSC should be
  13. 0:32carried out by one person at a time all
  14. 0:35work in a BSC should be carried out as
  15. 0:38far as possible towards the rear of the
  16. 0:40work area elbows and arms should not
  17. 0:43rest on the grill or work surface unless
  18. 0:46the BSC has been outfitted with
  19. 0:48appropriate arm rests one should avoid
  20. 0:51quick and excessive movement of hands
  21. 0:53and arms through the front opening such
  22. 0:55movements disrupt the air curtain at the
  23. 0:58front of the BSC
  24. 0:59which can allow contaminants to enter or
  25. 1:01escape arms should enter and exit the
  26. 1:04BSC slowly and perpendicular to the
  27. 1:07front opening a bottle of an appropriate
  28. 1:10disinfectant should be kept within the
  29. 1:12BSC while work is being performed this
  30. 1:16will reduce the need to remove your
  31. 1:17hands from the BSC it is recommended to
  32. 1:21segregate non contaminated items also
  33. 1:24known as clean items from contaminated
  34. 1:27items also known as dirty items work
  35. 1:30should always flow from clean to dirty
  36. 1:32areas material should be discarded in a
  37. 1:36waste container that is located towards
  38. 1:38the side of the cabinet workspace do not
  39. 1:42discard materials in containers outside
  40. 1:44of the cabinet open flames such as
  41. 1:47Bunsen burners should not be used inside
  42. 1:50a BSC as they create turbulence disrupt
  43. 1:53airflow patterns and can damage the HEPA
  44. 1:56filter if absolutely necessary touch
  45. 2:00activated micro burners are preferred to
  46. 2:02reduce airflow disturbance non flame
  47. 2:06alternatives such as sterile disposable
  48. 2:08inoculation loops
  49. 2:10be used whenever possible equipment
  50. 2:13creating air movement such as vacuum
  51. 2:15pumps and centrifuges may affect the
  52. 2:18integrity of the airflow and should not
  53. 2:20be used within the BSC and finally
  54. 2:23windows should be kept closed and
  55. 2:25cooling fans should be turned off when
  56. 2:27the BSC is in use in order to minimize
  57. 2:30disruptions to the air movement once you
  58. 2:33have completed your work you still have
  59. 2:35a few additional procedures to follow
  60. 2:37before turning off the BSC containers
  61. 2:41should be closed or covered and all
  62. 2:43items should be surface decontaminated
  63. 2:46before removing them from the BSC
  64. 2:48disinfect the interior surfaces of the
  65. 2:51BSC including sides back and interior of
  66. 2:55the glass use a disinfectant that is
  67. 2:58effective against the pathogens in use
  68. 3:00allowing an appropriate contact time if
  69. 3:03a corrosive disinfectant is used the
  70. 3:06surface should be wiped down after
  71. 3:08disinfection and sprayed with 70%
  72. 3:10ethanol to avoid corrosion of the
  73. 3:13stainless steel surfaces one should also
  74. 3:16routinely remove the work surface and
  75. 3:18disinfect the tray beneath you should
  76. 3:21also allow sufficient time for the air
  77. 3:23in the BSC to purge before disrupting
  78. 3:26the air curtain by removing hands or
  79. 3:28unloading material from the BSC the
  80. 3:31purge time will vary by model and can be
  81. 3:34up to several minutes the use of
  82. 3:36ultraviolet UV germicidal lamps to
  83. 3:39disinfect a contaminated BSC is strongly
  84. 3:42discouraged due to their limited
  85. 3:44effectiveness at disinfecting the inside
  86. 3:47of BSC s UV irradiation of the work area
  87. 3:51should only be used as a secondary
  88. 3:53method of disinfection in the cabinet
  89. 3:56let's conclude this video with a review
  90. 3:59of the following key messages a facility
  91. 4:02standard operating procedures should
  92. 4:04include best practices that minimize the
  93. 4:06disruption of air flow while working in
  94. 4:09a BSC
  95. 4:10work from clean to dirty and dispose of
  96. 4:13contaminated waste inside the cabinet it
  97. 4:16is important to decontaminate items and
  98. 4:19surfaces within the BSC after completing
  99. 4:22your work do not rely on UV irradiation
  100. 4:26alone to disinfect a contaminated work
  101. 4:29area knowledge of how to work properly
  102. 4:32in a BSC is a key component to working
  103. 4:35safely when using them
  104. 4:36[Music]
  105. 4:38you

About this transcript

This page contains the full transcript of Biological safety cabinet (BSC) 3: Best practices for safe usage by World Health Organization (WHO), generated from the public captions YouTube serves with the video. The transcript has 651 words across 105 segments, with the original timestamps preserved so you can click any line to jump to that moment in the embedded player.

What you can do with it

Use the transcript to take notes, quote the speaker, build a study guide, generate a summary with ChatGPT or Claude via the YouTube Summary tool, or export it as a timed subtitle file with YouTube to SRT. You can also re-open it in the transcriber to translate the transcript into 100+ languages.

Free YouTube transcript tool

YouTube2Text is a free YouTube transcript generator — no signup, no daily limit. Paste any YouTube link and get the full transcript instantly, with timestamps, click-to-jump, translation to 100+ languages, AI prompts for ChatGPT, Claude, and Gemini, and exports to TXT, SRT, VTT, or Markdown.