Biological safety cabinet (BSC) 1: Introduction — Transcript
Full transcript
- 0:04biological safety cabinets or BSC's are
- 0:07designed to protect laboratory workers
- 0:09and the surrounding environment from
- 0:11pathogens it is important that everyone
- 0:14working in a BSc understands when they
- 0:17should be used how they work and how to
- 0:19use them correctly BSC's provide
- 0:22effective primary containment for work
- 0:24with infectious material if they are
- 0:26properly maintained and used in
- 0:28combination with good laboratory
- 0:30techniques a risk assessment should be
- 0:32performed to determine whether a BSC is
- 0:35required for each activity conducted in
- 0:37your facility in general a BSC should be
- 0:41used when the activities increase the
- 0:43likelihood of exposure to pathogens and
- 0:45when the consequences of such an
- 0:47exposure could be serious to either the
- 0:50lab personnel or the environment for
- 0:52example a BSC can significantly reduce
- 0:55the risk when handling open vessels of
- 0:57pathogens transmitted via the inhalation
- 0:59route and when manipulating large
- 1:02volumes or highly concentrated
- 1:03infectious agents personnel protection
- 1:06is provided by BSC's through a
- 1:08continuous stream of inward air known as
- 1:11inflow which helps prevent aerosols from
- 1:13escaping through the front opening
- 1:16contaminated air is exhausted through
- 1:18high-efficiency particulate air filters
- 1:20or hepa filters before being released in
- 1:23the environment certain types of BSC s
- 1:26also protect the samples from becoming
- 1:28contaminated when working with them
- 1:30these cabinets use HEPA filtered down
- 1:33flow air to flush the cabinet interior
- 1:35of airborne contaminants and to prevent
- 1:37unfiltered inflow air from entering the
- 1:40work area it is important not to confuse
- 1:43BSC s with other ventilated enclosures
- 1:46commonly found in laboratories such as
- 1:48fume hoods and laminar flow cabinets
- 1:51which should never be used for personal
- 1:53protection against working with samples
- 1:55containing pathogens
- 1:57there are many classes and types of BS
- 1:59CS that offer varying levels of
- 2:01protection class 1b SCS provide
- 2:05personnel and environmental protection
- 2:07by drawing-room air into the cabinet
- 2:09through the front opening across the
- 2:11workspace and then discharging it from
- 2:13the cabinet through a HEPA filter the
- 2:16filtered exhaust air can be recirculated
- 2:18into the laboratory or exhausted
- 2:20directly to the outside atmosphere if it
- 2:22is hard ducted to the facility's HVAC
- 2:25system it is possible to work safely
- 2:28with small quantities of volatile toxic
- 2:30chemicals
- 2:31if the BSC is hard ducted class 1b SCS
- 2:35are generally recommended for procedures
- 2:37where total sterility of the biological
- 2:39material is not a critical factor these
- 2:42cabinets are often used for disposal of
- 2:45infectious waste placing samples in
- 2:47disinfectant and cage dumping or
- 2:49activities where infectious material
- 2:52won't be open but the risks of aerosol
- 2:54generation in the event of a malfunction
- 2:56are high homogeneous ation is one
- 2:59example of a process ideally suited for
- 3:02a class 1 BSC however in situations
- 3:05where no class 2 BSC is available there
- 3:09are other scenarios where class 1b SCS
- 3:12can be considered despite the lack of
- 3:14sample protection these include
- 3:16processes were the risk to personnel in
- 3:18the environment outweigh the need for
- 3:20sterility preparation of tissue sections
- 3:23for formal and fixation mechanical
- 3:26disruption of tissue with mortar and
- 3:28pestle addition of liquid samples or
- 3:30homogenates to inactivating agents for
- 3:33downstream nucleic acid or protein
- 3:35purification and preparation of smears
- 3:38for tuberculosis or rabies diagnosis are
- 3:41a few examples where even without
- 3:43product protection the likelihood of a
- 3:45negatively impacted end product or test
- 3:47result is low other activities such as
- 3:50cell culture virus isolation or
- 3:53manipulation of extracted nucleic acids
- 3:56should never be performed in a class 1
- 3:59BSC since the unfiltered air blowing
- 4:02over the work surface is likely to
- 4:03result in contamination loss of sample
- 4:06integrity and ultimately unreliable
- 4:09results class to be SCS provide
- 4:12personnel and environmental protection
- 4:15however unlike class 1b SCS
- 4:18they also offer product protection by
- 4:20filtering all air that passes through
- 4:22the front grills before distributing it
- 4:24over the work surface class 2b SCS are
- 4:27further divided into four types a 1 a 2
- 4:31B 1 and B 2 which differ according to in
- 4:36flow velocities recycling of cabinet air
- 4:38and connections to building exhaust
- 4:40systems class 2 type a1 and a2 cabinets
- 4:45are the most commonly encountered BSC s
- 4:47in microbiology Laboratories these
- 4:50cabinets recirculate 70% of HEPA
- 4:53filtered air flow back into the cabinet
- 4:55and exhaust the remaining air out of the
- 4:57cabinet through a second HEPA filter
- 4:59directly into the room class 2 type B
- 5:02cabinets differ from type a in they're
- 5:04higher in flow velocities and direct
- 5:06connection to the building's exhaust
- 5:08system class 2b ones recycle about 50%
- 5:12of HEPA filtered air back into the
- 5:14cabinet therefore work with only low
- 5:16levels of volatile toxic chemicals and
- 5:19trace amounts of radionuclides may be
- 5:21performed class 2b 2's discharge 100% of
- 5:25HEPA filtered cabinet air out through
- 5:27the building's exhaust system and
- 5:29therefore work with volatile toxic
- 5:31chemicals and radionuclides may be
- 5:33performed while these cabinets provide
- 5:36excellent levels of protection they have
- 5:38higher electrical consumption more
- 5:40complex installation requirements and
- 5:42increased overall lifetime running costs
- 5:45compared to type a cabinets all of these
- 5:48factors should be considered when a
- 5:50facility is preparing to purchase a BSC
- 5:52class 3 BS CS provide the maximum level
- 5:56of personnel and environmental
- 5:58protection they are completely enclosed
- 6:00and are kept under negative pressure
- 6:02these highly complex containment devices
- 6:05are typically reserved for pathogens and
- 6:07procedures where the consequence of a
- 6:09laboratory exposure or environmental
- 6:12release could be severe due to high
- 6:14associated more tau
- 6:15and transmissibility combined with a
- 6:17lack of effective treatment or
- 6:19prophylaxis all biosafety cabinets must
- 6:23be certified by manufacturers for
- 6:25compliance with a set of performance
- 6:26standards for BSC s these can vary
- 6:29depending on the manufacturer's country
- 6:32of origin but are typically NSF
- 6:34international or the European Committee
- 6:36for Standardization ideally BSC should
- 6:40also be certified on-site after
- 6:42installation and on an annual basis to
- 6:45ensure that supply and exhaust HEPA
- 6:47filters are functioning correctly and
- 6:49directional air flow is being maintained
- 6:52finally careful consideration must be
- 6:55given to where the BSC is installed
- 6:57within the room as a general rule select
- 7:01a location where minimal disturbance to
- 7:03air flow will occur to ensure its
- 7:05maximum effectiveness
- 7:06BSC s should not be located in
- 7:09high-traffic areas or directly opposite
- 7:11of seated workstations and a minimum
- 7:14distance must be maintained from other
- 7:16BSC's or chemical fume hoods to avoid
- 7:19disruption of the air curtain a local
- 7:21risk assessment will help determine a
- 7:23reasonably safe distance to avoid
- 7:25operation collision let us conclude this
- 7:29video with a review of the following key
- 7:31messages BSC's provide effective primary
- 7:35containment for work with infectious
- 7:37material if they are properly maintained
- 7:39and used in combination with good
- 7:41laboratory techniques there are many
- 7:43classes and types of BSC s that offer
- 7:46varying levels of protection a risk
- 7:48assessment will help determine which one
- 7:50is best suited for your laboratory and
- 7:52for the types of activities you will be
- 7:54conducting BSC s should be installed in
- 7:57areas with minimal foot traffic and
- 7:59potential airflow disturbances regular
- 8:02on-site recertification ensures proper
- 8:05functioning critical components
- 8:06including air flow profiles and hepa
- 8:08filters everyone working in a BSC should
- 8:12understand how they work when they
- 8:14should be used and how to use them
- 8:16correctly
- 8:17knowledge of how a BSc works is a key
- 8:20component to working safely when using
- 8:22them
- 8:25you
About this transcript
This page contains the full transcript of Biological safety cabinet (BSC) 1: Introduction by World Health Organization (WHO), generated from the public captions YouTube serves with the video. The transcript has 1,206 words across 205 segments, with the original timestamps preserved so you can click any line to jump to that moment in the embedded player.
What you can do with it
Use the transcript to take notes, quote the speaker, build a study guide, generate a summary with ChatGPT or Claude via the YouTube Summary tool, or export it as a timed subtitle file with YouTube to SRT. You can also re-open it in the transcriber to translate the transcript into 100+ languages.
Free YouTube transcript tool
YouTube2Text is a free YouTube transcript generator — no signup, no daily limit. Paste any YouTube link and get the full transcript instantly, with timestamps, click-to-jump, translation to 100+ languages, AI prompts for ChatGPT, Claude, and Gemini, and exports to TXT, SRT, VTT, or Markdown.