Atomic Absorption Spectroscopy Part 1 — Transcript
Full transcript
- 0:01the first thing we're going to do is
- 0:03turn on the um the air that we're going
- 0:07to use to uh uh provide oxygen for the
- 0:11flame this is the acetylene right
- 0:15here okay and uh what we're going to do
- 0:18is record the initial ESI it should
- 0:21never go below 100 uh so just to get an
- 0:25idea of usage we record his pressure at
- 0:27the beginning and then at the end should
- 0:30never go below 500 so it's 1,000 right
- 0:32now and this one 250 so we're in good
- 0:35shape we should record it so this is the
- 0:38atomic absorption spectrometer there's a
- 0:42long wedge and it's essentially where
- 0:44the flame is going to be it's a
- 0:46elongated flame and the flame consists
- 0:50of the air and the acetylene that we uh
- 0:54showed you and the lame is the flame is
- 0:57going to be lit and it's going to
- 0:59actually uh suck up uh the solution in
- 1:03this case it's just di water but it's
- 1:06going to suck up our sample Solutions
- 1:08and calibration Solutions into that
- 1:10flame and it become um lead atoms so
- 1:16we've turned on the computer and um we
- 1:19have to turn on the wall switch this the
- 1:22switch yellow tab and essentially what
- 1:26it does it starts with the fan that
- 1:29takes the exhaust
- 1:30from the flame up to the roof and out
- 1:33and machine will not turn on without
- 1:36that then we're going to uh turn on the
- 1:40instrument you're going to hear a
- 1:41grinding noise soon and that grinding
- 1:44noise is a good
- 1:46thing there it is okay what that noise
- 1:49is is the different wavelengths are
- 1:54accessed through a very fine def
- 1:56defraction grading and this grading
- 1:59essentially was not kind of finding its
- 2:01home place and that's what the grinding
- 2:04noise
- 2:07was we've turned on the instrument it's
- 2:11being warmed up it's got the gas that it
- 2:13needs and we're going to start the start
- 2:15the
- 2:18prog and we're going to pick up a
- 2:20worksheet that we've used
- 2:23before uh from
- 2:26lead and uh we're going to use that as
- 2:29our model cuz a lot of the same settings
- 2:31are
- 2:33there so we're going to say okay give it
- 2:37a name so the name is going to be Kem
- 2:42216 and for example um Monday would be
- 2:46the class okay and so we say
- 2:49okay then as uh we get a um since this
- 2:55has been a template we we've got it set
- 2:58up we're going to change it
- 3:00and U to do that um we're going to go to
- 3:06develop we're going to develop a method
- 3:08but look here since I selected the lead
- 3:11method that existed before I'm going to
- 3:13open this up and you can see that one of
- 3:15the lamps is lit and that's a a lamp
- 3:19that
- 3:19contains uh lead so in this lamp it's
- 3:23sending out um um emission from excited
- 3:28lead atoms so this lamp in it inside
- 3:31here has excited lead atoms that are
- 3:33spitting out just the right wavelength
- 3:37uh for lead atoms only and it's going
- 3:39through here it will go through the
- 3:42flame and if our sample has more or less
- 3:45lead in it the atoms that are in the uh
- 3:48flame will absorb that light coming from
- 3:50the lamp and decrease the signal and
- 3:52that's the basis um while the lamp is
- 3:55warming up it has to warm up for about
- 3:58uh about 5 or 10 10 minutes so we're
- 4:01going to edit the method that was
- 4:03existing and there are a lot of tabs
- 4:05that you see here so y we're doing lead
- 4:08it's in 2% nitric acid because we've
- 4:11made our standards that way and our
- 4:13samples uh are were adjusted by adding a
- 4:17nitric acid we're in the sample manual
- 4:19mode we're measuring absorption you can
- 4:22actually measure some elements just by
- 4:24the amount of light that they emit uh if
- 4:28they're strong emitters in this case
- 4:30lead uh emits emission is not very
- 4:34strong so we're going to use the lamp
- 4:36and measure the decrease in the light
- 4:38intensity by the
- 4:40detector so this is our air settling
- 4:43flame we're going to leave that the same
- 4:46uh in terms of measurement where you can
- 4:49measure it very different ways and this
- 4:51is perhaps for more Advanced Analytical
- 4:53course where we get into it but there
- 4:55are different ways to um take the data
- 4:58the absorption over period of time to
- 5:00get different kinds of averages we're
- 5:01going to integrate the um amount of
- 5:07absorption over a certain time period so
- 5:09we're going to use that minimum reading
- 5:11zero we're going to use some smoothing
- 5:14again some Advanced Techniques and each
- 5:17measurement is going to take 5 seconds
- 5:20so it takes a lot longer to make these
- 5:21samples than it is to actually measure
- 5:24them and we have a delay of of of six
- 5:27seconds so when you put this into your
- 5:31into your
- 5:32sample uh you have to clear out the old
- 5:34solution and then get the new solution
- 5:37coming in that you're measuring it and
- 5:38we're saying we're giving it about 6
- 5:40seconds which should be fine um we're
- 5:42going to take three replicates same
- 5:44sample but we're measuring at three
- 5:46times to get an average and we're
- 5:48measuring
- 5:49concentration kind of what we call an
- 5:52external calibration technique we're
- 5:54going to be looking at a a standard
- 5:56additions method uh in the next LA
- 5:59that's a little bit
- 6:01different um in terms of optical it's a
- 6:04land position three and the wavelength
- 6:07that we're going to be choosing to
- 6:10measure the quantity of the lead atoms
- 6:13is at 2177 NM so it's way way into the
- 6:16into the
- 6:18uh into the ultraviolet so we can't
- 6:20really see that and in order to uh
- 6:25provide um kind of a background
- 6:28correction we have another lamp that you
- 6:31don't see that's internal here which is
- 6:33a detarium lamp that puts out a lot of
- 6:36light at around
- 6:38217 that can compensate for other
- 6:41species that might be also absorbing at
- 6:44217 so here's where we put our standards
- 6:46so what standards do we have today
- 6:502.5 one two and five so we're not going
- 6:54to do the
- 6:5610
- 6:5730 so that where we put our standard
- 7:00concentration and in terms of
- 7:03calibration this says what kind of curve
- 7:05do you want to draw is it going to be a
- 7:07linear regression is it going to be what
- 7:09kind of fit are you going to use we're
- 7:12going to use something that's
- 7:13particularly well fitted to Atomic
- 7:16absorption spectrometry which has a kind
- 7:19of
- 7:20a follows Beer's Law pretty well but uh
- 7:24if the congration gets too high uh the
- 7:27absorption kind of plant and so for this
- 7:31instrument there's a new kind of a a
- 7:33parameter called new rational and we're
- 7:36going to use that it's going to fit the
- 7:37data for us so we don't have to actually
- 7:39plot the data the next is
- 7:43notes we're going to uh use a is just
- 7:47notes if you're doing an experiment to
- 7:49remind you to do this and that this is a
- 7:51fun feature this is called The cookbook
- 7:54The cookbook for each element that
- 7:56you're looking at you zoom in here you
- 7:59can uh uh show you that there are
- 8:00different wavelengths that you can
- 8:02analyze lead for if you're if you have a
- 8:05concentration that has a heck of a lot
- 8:07of lead then you don't need a very
- 8:10intense uh uh wavelength you can use
- 8:13light of less wavelength so we're going
- 8:17to be using wavelength of of where it's
- 8:21the most
- 8:23sensitive um and we expect for2
- 8:25milligrams per liter um for for five
- 8:30milligrams of uh per liter an absorption
- 8:33of about 0.2 and that's kind of a good
- 8:36re region to be in it also tells you
- 8:39what kind of absorbances you have so if
- 8:42you're analyzing something that um has
- 8:45uh for example um a lot of phosphate it
- 8:48tells you what you can do to you know
- 8:51get rid of that but we're just dealing
- 8:53with water and Lead sulfate so it
- 8:56shouldn't be a problem this is a graph
- 8:59which kind TR to predicts what your
- 9:01calibration Curve will look like and
- 9:04this is the
- 9:05concentration uh and so if you have
- 9:08absorption about five uh milligrams per
- 9:12liter or 5 PPM you should get an
- 9:14absorption of 02 at this waiver qcp this
- 9:17is quality control okay and we talked
- 9:20about this a lot so it allows you to uh
- 9:25run a blank every so often just to make
- 9:28sure that the zero is not shifting so
- 9:30I've I've clicked that on and um it's a
- 9:34note to self that the blank should not
- 9:38be higher than 0.1 otherwise I'll an
- 9:41error message so that's a good thing
- 9:43every so often we're going to do a
- 9:45continuing calibration verification
- 9:49that's going to be at 1 mgram per liter
- 9:52so we'll run this one every so often to
- 9:54make sure we get a reading of one part
- 9:57per million okay so that's another quity
- 9:59control and if it's between if it's
- 10:02between 80 and
- 10:03120% of uh one it'll say have a good day
- 10:08that's a good number if it's out of that
- 10:10it'll say it'll put a flag up and say
- 10:12you're out of the mpect and I've told it
- 10:14just to
- 10:15continue if you're a lab uh that's doing
- 10:19commercial work you would stop the
- 10:21process make sure you're inspect but I
- 10:23think we're going to be okay the lamp is
- 10:26warmed up now what we're going to do is
- 10:29we're going to put labels so these are
- 10:32the samples that we're going to be
- 10:34running and so the sample we're going to
- 10:36be running is the sample that you will
- 10:39have
- 10:40prepared um
- 10:42from uh your
- 10:45precipitated U Le s okay so we would
- 10:49have uh different initials we would have
- 10:51maybe m
- 10:55b okay okay
- 10:59D okay all all of the teams in the L
- 11:03notice here I have a continuing
- 11:06calibration but I for this
- 11:11particular see B notice it turns
- 11:16red
- 11:18CCV so now we' put in the samples we've
- 11:21put in the calibration and now we go to
- 11:24analysis so one of the things we want to
- 11:26do is we want to indicate where we're
- 11:30going to beting the data so we go to
- 11:35select um and we want to um erase those
- 11:40that we're not
- 11:42doing and come back so that'll be
- 11:47the samples that we're looking at here
- 11:50now we have to optimize the uh lamp now
- 11:54that it's been started and um
- 12:00so we
- 12:02say
- 12:03okay and you hear the machine makes a
- 12:06little dinging sound that's the
- 12:07monochrometer which is coming into the
- 12:10wavelength down here you can see peeking
- 12:12at lead
- 12:13217 that's the most sensitive wave for
- 12:17the
- 12:18lead click optimize uh lamps okay and if
- 12:23you see over here we have some green
- 12:25bars which indicates the intensity of
- 12:27the light
- 12:29uh this is the um HC light which stands
- 12:32for Hol cathode lamp that's the type of
- 12:35lamp that we have in here Hollow cathod
- 12:38lamps notice there's slots for other
- 12:40elements too so if you have a sample
- 12:42that has a number of elements you can do
- 12:44them with the same sample in can it is
- 12:46essentially excited lead atoms that's
- 12:48what they look like and that light is
- 12:50then going out through the monochrometer
- 12:52you can see right over
- 12:55here okay and then it's directed right
- 12:57through here and part of this setup is
- 13:00to make sure that you have that
- 13:02lamp uh kind of set up parallel so if
- 13:05you take a look over my
- 13:08shoulder you see this this this card has
- 13:12a Target area and you put this is right
- 13:15where the slot is
- 13:17and uh you put it right over there and
- 13:22you want to make sure that as you go
- 13:24back and forth it's it's essentially now
- 13:27why would you say why would it be
- 13:29changed well there's a little handle
- 13:31here if you're dealing with too high
- 13:33concentration you don't want much
- 13:35absorption you can you can actually go
- 13:37through the through the flame at an
- 13:39angle and get less absorption so it's
- 13:41very versatile uh this is a knob that
- 13:44lets you go up and down and uh sideways
- 13:47to fine to the position of the flame I'm
- 13:49now I'm going to take this now that I've
- 13:51got the position right this
- 13:55here this is a shield over here
- 14:00these lamps are putting out high
- 14:02intensity UV light but this is glass
- 14:04which absorbs those the very end of
- 14:07those tubes essentially are quch and
- 14:09will pass the UV light but it's safe to
- 14:11look at the lamps but not the flame and
- 14:14one of the things to note is the gain
- 14:17down here you'll write that in your
- 14:18notebook uh the uh gain should be um
- 14:24recorded because it represents the
- 14:26amount of amplification the electronics
- 14:28are doing as the gain goes higher and
- 14:31higher it me the lamp is getting weaker
- 14:33and weaker so you want to monitor that
- 14:36so um essentially optimize uh lamps
- 14:41notice that these two bars here are
- 14:43about the same height that is desirable
- 14:46we're going to
- 14:52find and we say
- 14:57okay cancel
- 14:59okay
About this transcript
This page contains the full transcript of Atomic Absorption Spectroscopy Part 1 by Quantitative Analysis, generated from the public captions YouTube serves with the video. The transcript has 2,240 words across 323 segments, with the original timestamps preserved so you can click any line to jump to that moment in the embedded player.
What you can do with it
Use the transcript to take notes, quote the speaker, build a study guide, generate a summary with ChatGPT or Claude via the YouTube Summary tool, or export it as a timed subtitle file with YouTube to SRT. You can also re-open it in the transcriber to choose a caption track available on the video.
Free YouTube transcript tool
YouTube2Text is a free YouTube transcript generator — no signup, no daily limit. Paste any YouTube link and get the full transcript instantly, with timestamps, click-to-jump, caption language selection, AI prompts for ChatGPT, Claude, and Gemini, and exports to TXT, SRT, VTT, or Markdown.