  1. 0:01the first thing we're going to do is
  2. 0:03turn on the um the air that we're going
  3. 0:07to use to uh uh provide oxygen for the
  4. 0:11flame this is the acetylene right
  5. 0:15here okay and uh what we're going to do
  6. 0:18is record the initial ESI it should
  7. 0:21never go below 100 uh so just to get an
  8. 0:25idea of usage we record his pressure at
  9. 0:27the beginning and then at the end should
  10. 0:30never go below 500 so it's 1,000 right
  11. 0:32now and this one 250 so we're in good
  12. 0:35shape we should record it so this is the
  13. 0:38atomic absorption spectrometer there's a
  14. 0:42long wedge and it's essentially where
  15. 0:44the flame is going to be it's a
  16. 0:46elongated flame and the flame consists
  17. 0:50of the air and the acetylene that we uh
  18. 0:54showed you and the lame is the flame is
  19. 0:57going to be lit and it's going to
  20. 0:59actually uh suck up uh the solution in
  21. 1:03this case it's just di water but it's
  22. 1:06going to suck up our sample Solutions
  23. 1:08and calibration Solutions into that
  24. 1:10flame and it become um lead atoms so
  25. 1:16we've turned on the computer and um we
  26. 1:19have to turn on the wall switch this the
  27. 1:22switch yellow tab and essentially what
  28. 1:26it does it starts with the fan that
  29. 1:29takes the exhaust
  30. 1:30from the flame up to the roof and out
  31. 1:33and machine will not turn on without
  32. 1:36that then we're going to uh turn on the
  33. 1:40instrument you're going to hear a
  34. 1:41grinding noise soon and that grinding
  35. 1:44noise is a good
  36. 1:46thing there it is okay what that noise
  37. 1:49is is the different wavelengths are
  38. 1:54accessed through a very fine def
  39. 1:56defraction grading and this grading
  40. 1:59essentially was not kind of finding its
  41. 2:01home place and that's what the grinding
  42. 2:04noise
  43. 2:07was we've turned on the instrument it's
  44. 2:11being warmed up it's got the gas that it
  45. 2:13needs and we're going to start the start
  46. 2:15the
  47. 2:18prog and we're going to pick up a
  48. 2:20worksheet that we've used
  49. 2:23before uh from
  50. 2:26lead and uh we're going to use that as
  51. 2:29our model cuz a lot of the same settings
  52. 2:31are
  53. 2:33there so we're going to say okay give it
  54. 2:37a name so the name is going to be Kem
  55. 2:42216 and for example um Monday would be
  56. 2:46the class okay and so we say
  57. 2:49okay then as uh we get a um since this
  58. 2:55has been a template we we've got it set
  59. 2:58up we're going to change it
  60. 3:00and U to do that um we're going to go to
  61. 3:06develop we're going to develop a method
  62. 3:08but look here since I selected the lead
  63. 3:11method that existed before I'm going to
  64. 3:13open this up and you can see that one of
  65. 3:15the lamps is lit and that's a a lamp
  66. 3:19that
  67. 3:19contains uh lead so in this lamp it's
  68. 3:23sending out um um emission from excited
  69. 3:28lead atoms so this lamp in it inside
  70. 3:31here has excited lead atoms that are
  71. 3:33spitting out just the right wavelength
  72. 3:37uh for lead atoms only and it's going
  73. 3:39through here it will go through the
  74. 3:42flame and if our sample has more or less
  75. 3:45lead in it the atoms that are in the uh
  76. 3:48flame will absorb that light coming from
  77. 3:50the lamp and decrease the signal and
  78. 3:52that's the basis um while the lamp is
  79. 3:55warming up it has to warm up for about
  80. 3:58uh about 5 or 10 10 minutes so we're
  81. 4:01going to edit the method that was
  82. 4:03existing and there are a lot of tabs
  83. 4:05that you see here so y we're doing lead
  84. 4:08it's in 2% nitric acid because we've
  85. 4:11made our standards that way and our
  86. 4:13samples uh are were adjusted by adding a
  87. 4:17nitric acid we're in the sample manual
  88. 4:19mode we're measuring absorption you can
  89. 4:22actually measure some elements just by
  90. 4:24the amount of light that they emit uh if
  91. 4:28they're strong emitters in this case
  92. 4:30lead uh emits emission is not very
  93. 4:34strong so we're going to use the lamp
  94. 4:36and measure the decrease in the light
  95. 4:38intensity by the
  96. 4:40detector so this is our air settling
  97. 4:43flame we're going to leave that the same
  98. 4:46uh in terms of measurement where you can
  99. 4:49measure it very different ways and this
  100. 4:51is perhaps for more Advanced Analytical
  101. 4:53course where we get into it but there
  102. 4:55are different ways to um take the data
  103. 4:58the absorption over period of time to
  104. 5:00get different kinds of averages we're
  105. 5:01going to integrate the um amount of
  106. 5:07absorption over a certain time period so
  107. 5:09we're going to use that minimum reading
  108. 5:11zero we're going to use some smoothing
  109. 5:14again some Advanced Techniques and each
  110. 5:17measurement is going to take 5 seconds
  111. 5:20so it takes a lot longer to make these
  112. 5:21samples than it is to actually measure
  113. 5:24them and we have a delay of of of six
  114. 5:27seconds so when you put this into your
  115. 5:31into your
  116. 5:32sample uh you have to clear out the old
  117. 5:34solution and then get the new solution
  118. 5:37coming in that you're measuring it and
  119. 5:38we're saying we're giving it about 6
  120. 5:40seconds which should be fine um we're
  121. 5:42going to take three replicates same
  122. 5:44sample but we're measuring at three
  123. 5:46times to get an average and we're
  124. 5:48measuring
  125. 5:49concentration kind of what we call an
  126. 5:52external calibration technique we're
  127. 5:54going to be looking at a a standard
  128. 5:56additions method uh in the next LA
  129. 5:59that's a little bit
  130. 6:01different um in terms of optical it's a
  131. 6:04land position three and the wavelength
  132. 6:07that we're going to be choosing to
  133. 6:10measure the quantity of the lead atoms
  134. 6:13is at 2177 NM so it's way way into the
  135. 6:16into the
  136. 6:18uh into the ultraviolet so we can't
  137. 6:20really see that and in order to uh
  138. 6:25provide um kind of a background
  139. 6:28correction we have another lamp that you
  140. 6:31don't see that's internal here which is
  141. 6:33a detarium lamp that puts out a lot of
  142. 6:36light at around
  143. 6:38217 that can compensate for other
  144. 6:41species that might be also absorbing at
  145. 6:44217 so here's where we put our standards
  146. 6:46so what standards do we have today
  147. 6:502.5 one two and five so we're not going
  148. 6:54to do the
  149. 6:5610
  150. 6:5730 so that where we put our standard
  151. 7:00concentration and in terms of
  152. 7:03calibration this says what kind of curve
  153. 7:05do you want to draw is it going to be a
  154. 7:07linear regression is it going to be what
  155. 7:09kind of fit are you going to use we're
  156. 7:12going to use something that's
  157. 7:13particularly well fitted to Atomic
  158. 7:16absorption spectrometry which has a kind
  159. 7:19of
  160. 7:20a follows Beer's Law pretty well but uh
  161. 7:24if the congration gets too high uh the
  162. 7:27absorption kind of plant and so for this
  163. 7:31instrument there's a new kind of a a
  164. 7:33parameter called new rational and we're
  165. 7:36going to use that it's going to fit the
  166. 7:37data for us so we don't have to actually
  167. 7:39plot the data the next is
  168. 7:43notes we're going to uh use a is just
  169. 7:47notes if you're doing an experiment to
  170. 7:49remind you to do this and that this is a
  171. 7:51fun feature this is called The cookbook
  172. 7:54The cookbook for each element that
  173. 7:56you're looking at you zoom in here you
  174. 7:59can uh uh show you that there are
  175. 8:00different wavelengths that you can
  176. 8:02analyze lead for if you're if you have a
  177. 8:05concentration that has a heck of a lot
  178. 8:07of lead then you don't need a very
  179. 8:10intense uh uh wavelength you can use
  180. 8:13light of less wavelength so we're going
  181. 8:17to be using wavelength of of where it's
  182. 8:21the most
  183. 8:23sensitive um and we expect for2
  184. 8:25milligrams per liter um for for five
  185. 8:30milligrams of uh per liter an absorption
  186. 8:33of about 0.2 and that's kind of a good
  187. 8:36re region to be in it also tells you
  188. 8:39what kind of absorbances you have so if
  189. 8:42you're analyzing something that um has
  190. 8:45uh for example um a lot of phosphate it
  191. 8:48tells you what you can do to you know
  192. 8:51get rid of that but we're just dealing
  193. 8:53with water and Lead sulfate so it
  194. 8:56shouldn't be a problem this is a graph
  195. 8:59which kind TR to predicts what your
  196. 9:01calibration Curve will look like and
  197. 9:04this is the
  198. 9:05concentration uh and so if you have
  199. 9:08absorption about five uh milligrams per
  200. 9:12liter or 5 PPM you should get an
  201. 9:14absorption of 02 at this waiver qcp this
  202. 9:17is quality control okay and we talked
  203. 9:20about this a lot so it allows you to uh
  204. 9:25run a blank every so often just to make
  205. 9:28sure that the zero is not shifting so
  206. 9:30I've I've clicked that on and um it's a
  207. 9:34note to self that the blank should not
  208. 9:38be higher than 0.1 otherwise I'll an
  209. 9:41error message so that's a good thing
  210. 9:43every so often we're going to do a
  211. 9:45continuing calibration verification
  212. 9:49that's going to be at 1 mgram per liter
  213. 9:52so we'll run this one every so often to
  214. 9:54make sure we get a reading of one part
  215. 9:57per million okay so that's another quity
  216. 9:59control and if it's between if it's
  217. 10:02between 80 and
  218. 10:03120% of uh one it'll say have a good day
  219. 10:08that's a good number if it's out of that
  220. 10:10it'll say it'll put a flag up and say
  221. 10:12you're out of the mpect and I've told it
  222. 10:14just to
  223. 10:15continue if you're a lab uh that's doing
  224. 10:19commercial work you would stop the
  225. 10:21process make sure you're inspect but I
  226. 10:23think we're going to be okay the lamp is
  227. 10:26warmed up now what we're going to do is
  228. 10:29we're going to put labels so these are
  229. 10:32the samples that we're going to be
  230. 10:34running and so the sample we're going to
  231. 10:36be running is the sample that you will
  232. 10:39have
  233. 10:40prepared um
  234. 10:42from uh your
  235. 10:45precipitated U Le s okay so we would
  236. 10:49have uh different initials we would have
  237. 10:51maybe m
  238. 10:55b okay okay
  239. 10:59D okay all all of the teams in the L
  240. 11:03notice here I have a continuing
  241. 11:06calibration but I for this
  242. 11:11particular see B notice it turns
  243. 11:16red
  244. 11:18CCV so now we' put in the samples we've
  245. 11:21put in the calibration and now we go to
  246. 11:24analysis so one of the things we want to
  247. 11:26do is we want to indicate where we're
  248. 11:30going to beting the data so we go to
  249. 11:35select um and we want to um erase those
  250. 11:40that we're not
  251. 11:42doing and come back so that'll be
  252. 11:47the samples that we're looking at here
  253. 11:50now we have to optimize the uh lamp now
  254. 11:54that it's been started and um
  255. 12:00so we
  256. 12:02say
  257. 12:03okay and you hear the machine makes a
  258. 12:06little dinging sound that's the
  259. 12:07monochrometer which is coming into the
  260. 12:10wavelength down here you can see peeking
  261. 12:12at lead
  262. 12:13217 that's the most sensitive wave for
  263. 12:17the
  264. 12:18lead click optimize uh lamps okay and if
  265. 12:23you see over here we have some green
  266. 12:25bars which indicates the intensity of
  267. 12:27the light
  268. 12:29uh this is the um HC light which stands
  269. 12:32for Hol cathode lamp that's the type of
  270. 12:35lamp that we have in here Hollow cathod
  271. 12:38lamps notice there's slots for other
  272. 12:40elements too so if you have a sample
  273. 12:42that has a number of elements you can do
  274. 12:44them with the same sample in can it is
  275. 12:46essentially excited lead atoms that's
  276. 12:48what they look like and that light is
  277. 12:50then going out through the monochrometer
  278. 12:52you can see right over
  279. 12:55here okay and then it's directed right
  280. 12:57through here and part of this setup is
  281. 13:00to make sure that you have that
  282. 13:02lamp uh kind of set up parallel so if
  283. 13:05you take a look over my
  284. 13:08shoulder you see this this this card has
  285. 13:12a Target area and you put this is right
  286. 13:15where the slot is
  287. 13:17and uh you put it right over there and
  288. 13:22you want to make sure that as you go
  289. 13:24back and forth it's it's essentially now
  290. 13:27why would you say why would it be
  291. 13:29changed well there's a little handle
  292. 13:31here if you're dealing with too high
  293. 13:33concentration you don't want much
  294. 13:35absorption you can you can actually go
  295. 13:37through the through the flame at an
  296. 13:39angle and get less absorption so it's
  297. 13:41very versatile uh this is a knob that
  298. 13:44lets you go up and down and uh sideways
  299. 13:47to fine to the position of the flame I'm
  300. 13:49now I'm going to take this now that I've
  301. 13:51got the position right this
  302. 13:55here this is a shield over here
  303. 14:00these lamps are putting out high
  304. 14:02intensity UV light but this is glass
  305. 14:04which absorbs those the very end of
  306. 14:07those tubes essentially are quch and
  307. 14:09will pass the UV light but it's safe to
  308. 14:11look at the lamps but not the flame and
  309. 14:14one of the things to note is the gain
  310. 14:17down here you'll write that in your
  311. 14:18notebook uh the uh gain should be um
  312. 14:24recorded because it represents the
  313. 14:26amount of amplification the electronics
  314. 14:28are doing as the gain goes higher and
  315. 14:31higher it me the lamp is getting weaker
  316. 14:33and weaker so you want to monitor that
  317. 14:36so um essentially optimize uh lamps
  318. 14:41notice that these two bars here are
  319. 14:43about the same height that is desirable
  320. 14:46we're going to
  321. 14:52find and we say
  322. 14:57okay cancel
  323. 14:59okay

About this transcript

This page contains the full transcript of Atomic Absorption Spectroscopy Part 1 by Quantitative Analysis, generated from the public captions YouTube serves with the video. The transcript has 2,240 words across 323 segments, with the original timestamps preserved so you can click any line to jump to that moment in the embedded player.

What you can do with it

Use the transcript to take notes, quote the speaker, build a study guide, generate a summary with ChatGPT or Claude via the YouTube Summary tool, or export it as a timed subtitle file with YouTube to SRT. You can also re-open it in the transcriber to choose a caption track available on the video.

Free YouTube transcript tool

YouTube2Text is a free YouTube transcript generator — no signup, no daily limit. Paste any YouTube link and get the full transcript instantly, with timestamps, click-to-jump, caption language selection, AI prompts for ChatGPT, Claude, and Gemini, and exports to TXT, SRT, VTT, or Markdown.